J-TIP SYRINGE J-TIP 0.25 ML 8164-2001-25

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2013-11-20 for J-TIP SYRINGE J-TIP 0.25 ML 8164-2001-25 manufactured by National Medical Products, Inc..

Event Text Entries

[4166098] User reported malfunction, breakage or device (s) at use. (b)(4). "event desc: j-tip "exploded" while being administered prior to peripheral iv placement. Part of the j-tip closes to end shattered into mall pieces. Medication did not work as intended. Defective j-tip given to ed pharmacist. " not normal practice but cannot report that the devices were not stored outside labeling storage conditions. Syringe broke apart during activation. Pharmacy lot number.
Patient Sequence No: 1, Text Type: D, B5


[11489641] Manufacturer has tested units and found extreme heat (+ 140 degrees c) and extreme cooling (dry ice or freezer storage) can cause device to crack. These extreme temperatures are the only way the manufacturer could find cracking damage. No actual "breakage" occurred during testing but cracking of the main body of the device was observed. Attempt has been made to obtain device. If device is returned a follow-up will be filed with fda.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2027063-2013-00001
MDR Report Key3599280
Report Source06
Date Received2013-11-20
Date of Report2013-11-21
Date of Event2013-05-21
Date Mfgr Received2013-10-28
Date Added to Maude2014-03-12
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer Street57 PARKER ST.
Manufacturer CityIRVINE CA 92618
Manufacturer CountryUS
Manufacturer Postal92618
Manufacturer Phone9492853517
Single Use3
Remedial ActionOT
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameJ-TIP SYRINGE
Generic NameSYRINGE, NEEDLESS
Product CodeKZE
Date Received2013-11-20
Model NumberJ-TIP 0.25 ML
Catalog Number8164-2001-25
Lot NumberUNK
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerNATIONAL MEDICAL PRODUCTS, INC.
Manufacturer Address57 PARKER ST. IRVINE CA 92618 US 92618


Patients

Patient NumberTreatmentOutcomeDate
10 2013-11-20

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.