MINIVIDAS W2011

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01 report with the FDA on 2001-11-12 for MINIVIDAS W2011 manufactured by Biomerieux, Inc..

Event Text Entries

[221143] Customer was running "qct" test strips on pump and failed the test 4 times. Qcr test is a test to detect clogged pump tubes. Clogged pump tubes could result in false results being reported to the pt. There are no reports of false results being generated at this site.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1950204-2001-00016
MDR Report Key360507
Report Source01
Date Received2001-11-12
Date of Report2001-11-12
Date of Event2001-10-08
Date Mfgr Received2001-10-08
Device Manufacturer Date1994-12-01
Date Added to Maude2001-11-20
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location0
Manufacturer ContactSANDRA PERREAND
Manufacturer Street595 ANGLUM DRIVE
Manufacturer CityHAZELWOOD MO 63042
Manufacturer CountryUS
Manufacturer Postal63042
Manufacturer Phone*
Manufacturer G1*
Manufacturer Street*
Manufacturer City*
Manufacturer Country*
Single Use3
Remedial ActionRL
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameMINIVIDAS
Generic NameIMMUNOFLUROMETRIC EQUIPMENT
Product CodeIZT
Date Received2001-11-12
Model NumberMINIVIDAS
Catalog NumberW2011
Lot NumberNA
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Implant FlagN
Date RemovedA
Device Sequence No1
Device Event Key349676
ManufacturerBIOMERIEUX, INC.
Manufacturer Address595 ANGLUM DR. HAZELWOOD MO 63042 US
Baseline Brand NameMINIVIDAS
Baseline Generic NameIMMUNOFLUROMETRIC EQUIPMENT
Baseline Model NoMINIVIDAS
Baseline Catalog NoW2010
Baseline IDNA


Patients

Patient NumberTreatmentOutcomeDate
10 2001-11-12

© 2025 FDA.report
This site is not affiliated with or endorsed by the FDA.