CORE IMPACTION DRILL 5400300000

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2014-02-07 for CORE IMPACTION DRILL 5400300000 manufactured by Stryker Instruments-kalamazoo.

Event Text Entries

[4093520] It was reported that during testing conducted at the user facility the core impaction drill was overheating. No medical intervention and no adverse consequences were reported with this event. As this event occurred during testing at the user facility, there was no patient involvement and no delay to a surgical procedure.
Patient Sequence No: 1, Text Type: D, B5


[11660254] Upon disassembly for visual inspection corrosion was found within the device, the motor pins were damaged, a bearing was stuck on the rotor, the driveshaft was sticking, and the rotor coupler and spindle housing were worn.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number0001811755-2014-00394
MDR Report Key3612555
Report Source06
Date Received2014-02-07
Date of Report2014-01-13
Date of Event2014-01-09
Date Mfgr Received2014-01-13
Device Manufacturer Date2005-11-16
Date Added to Maude2014-04-30
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMR. CASEY METZGER
Manufacturer Street4100 EAST MILHAM AVENUE
Manufacturer CityKALAMAZOO MI 49001
Manufacturer CountryUS
Manufacturer Postal49001
Manufacturer Phone2693237700
Manufacturer G1STRYKER INSTRUMENTS-KALAMAZOO
Manufacturer Street4100 EAST MILHAM AVENUE
Manufacturer CityKALAMAZOO MI 49001
Manufacturer CountryUS
Manufacturer Postal Code49001
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCORE IMPACTION DRILL
Generic NameDRIVER, WIRE, AND BONE DRILL, MANUAL
Product CodeDZK
Date Received2014-02-07
Returned To Mfg2014-01-23
Catalog Number5400300000
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSTRYKER INSTRUMENTS-KALAMAZOO
Manufacturer Address4100 EAST MILHAM AVENUE KALAMAZOO MI 49001 US 49001


Patients

Patient NumberTreatmentOutcomeDate
10 2014-02-07