COBAS 6000 E601 MODULE 04745922001

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,05,06 report with the FDA on 2014-02-11 for COBAS 6000 E601 MODULE 04745922001 manufactured by Roche Diagnostics.

Event Text Entries

[4134710] The customer received a questionable antibody to hepatitis c virus (anti-hcv) result for one patient. The result from the cobas e 601 analyzer was 0. 459 coi (non-reactive). On (b)(6) 2014, the result from the abbott architect i1000 analyzer was 2. 93 s/co(positive). The result of 0. 459 coi was reported outside the laboratory. The patient was not adversely affected. The anti-hcv reagent lot number was 175153 with an expiration date of 06/29/2014.
Patient Sequence No: 1, Text Type: D, B5


[11624465] The investigation could not determine a specific root cause. Additional information for further investigation was requested, but was not provided. It was also noted that no patient sample could be provided for further investigation.
Patient Sequence No: 1, Text Type: N, H10


[11643668] This event occurred in (b)(6).
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1823260-2014-00902
MDR Report Key3622930
Report Source01,05,06
Date Received2014-02-11
Date of Report2014-02-26
Date of Event2014-01-25
Date Mfgr Received2014-01-28
Date Added to Maude2014-02-11
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1HITACHI HIGH TECH CORP.
Manufacturer Street882 ICHIGE HITACHINAKA NA
Manufacturer CityIBARAKI 312-8504
Manufacturer CountryJA
Manufacturer Postal Code312-8504
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameCOBAS 6000 E601 MODULE
Generic NameIMMUNOCHEMISTRY ANALYZER
Product CodeMZP
Date Received2014-02-11
Model NumberNA
Catalog Number04745922001
Lot NumberNA
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 46250 US 46250


Patients

Patient NumberTreatmentOutcomeDate
10 2014-02-11

© 2025 FDA.report
This site is not affiliated with or endorsed by the FDA.