VERITI DX THERMAL CYCLER 4452300

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 04,07 report with the FDA on 2013-12-09 for VERITI DX THERMAL CYCLER 4452300 manufactured by Life Technologies Holdings Pte. Ltd..

Event Text Entries

[16685769] A customer reported that the instrument veriti dx thermal cycler (catalogue no. 4452300) with serial no. (b)(4) was having a 'error - set point failure' message. No patient involvement reported.
Patient Sequence No: 1, Text Type: D, B5


[16711974] Device intended use: the applied biosystems veriti dx thermal cycler amplifies human nucleic acid samples for diagnostic applications. The veriti dx thermal cycler is to be used only by operators trained in laboratory techniques and procedures. Based on investigation conducted, the instrument gives an error once it reached 95 degrees c. It was tested with working controller pca and it worked. The repair was completed by replacing the controller pca. This is the initial and final report for this issue.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3003673482-2013-00008
MDR Report Key3623818
Report Source04,07
Date Received2013-12-09
Date of Report2013-05-20
Date of Event2013-05-01
Date Mfgr Received2013-05-20
Device Manufacturer Date2012-07-01
Date Added to Maude2014-03-31
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactKELLI TANZELLA
Manufacturer Street3175 STALEY ROAD
Manufacturer CityGRAND ISLAND NY 14072
Manufacturer CountryUS
Manufacturer Postal14072
Manufacturer Phone7167743122
Manufacturer G1LIFE TECHNOLOGIES HOLDINGS PTE. LTD.
Manufacturer StreetBLK 33, MARSILING INDUSTRIAL STATE ROAD 3, #07-06
Manufacturer CitySINGAPORE
Manufacturer CountrySN
Single Use3
Remedial ActionRP
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameVERITI DX THERMAL CYCLER
Generic NameOUL 862.2170
Product CodeNSU
Date Received2013-12-09
Catalog Number4452300
Lot Number299120182
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerLIFE TECHNOLOGIES HOLDINGS PTE. LTD.
Manufacturer AddressSINGAPORE SN


Patients

Patient NumberTreatmentOutcomeDate
10 2013-12-09