AFFIRM VPIII AMBIENT TEMPERATURE TRANSPORT 446255

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2014-02-05 for AFFIRM VPIII AMBIENT TEMPERATURE TRANSPORT 446255 manufactured by Becton, Dickinson.

Event Text Entries

[16803700] Customer reported a minor injury while running a pt sample on the affirm assay. A tech punctured her finger on a glass shard inside the specimen tube. The wound required basic first aid and the tech will received baseline testing.
Patient Sequence No: 1, Text Type: D, B5


[17048302] The affirm atts collection kit is used to stabilize the nucleic acid of candida species, gardnerella vaginalis, and trichomonas vaginalis during specimen transport at ambient temperature (15-30 degrees celsius) or (2-8 degrees celsius) for up to 72h. Bd determined that in this situation that the customer did not follow the testing procedure. Per technical svs, the customer opened the tip of the plastic dropper when dispensing the atts solution into the sample tube. This allowed the glass to be poured into the sample tube. (b)(4) no corrective action will be initiated at this time. Bd will continue to monitor for trends.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1119779-2014-00008
MDR Report Key3625097
Report Source05
Date Received2014-02-05
Date of Report2014-01-17
Date of Event2014-01-17
Date Mfgr Received2014-01-17
Date Added to Maude2014-02-14
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactCHARLOTTE DANNENFELSER
Manufacturer Street7 LOVETON CIR.
Manufacturer CitySPARKS MD 21152
Manufacturer CountryUS
Manufacturer Postal21152
Manufacturer Phone4103164367
Manufacturer G1BECTON DICKINSON
Manufacturer StreetPERIFERICO LUIS DONALDO, COLODIO #579
Manufacturer CityNOGALES, SONORA, C.P. 84000
Manufacturer CountryMX
Manufacturer Postal Code84000
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameAFFIRM VPIII AMBIENT TEMPERATURE TRANSPORT
Product CodeLIO
Date Received2014-02-05
Catalog Number446255
Lot Number3038199
Device Expiration Date2014-08-13
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBECTON, DICKINSON
Manufacturer AddressNOGALES, SONORA, C.P. MX


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2014-02-05

© 2024 FDA.report