BREMER HALO SYSTEM PT002

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 08 report with the FDA on 2001-12-05 for BREMER HALO SYSTEM PT002 manufactured by Depuy Acromed, Inc..

Event Text Entries

[236877] It was reported to depuy acromed that the securing threaded knobs on the lateral portion of the halo were broken, causing the halo to loosen. According to the complaintant, this break occurred eleven days after initial application. The patient's spine was stabiled and another halo was applied. A complaint history was performed and no other complaints have been reported for this part number. A dimensional analysis was performed and the halo conformed to specifications. And exact cause of the event cannot be determined, it is possible that the halo securing screw could have been over-torqued during application, causing the knob to fracture. The material of the threaded securing knobs is being changed from plastic to aluminum. This will increase the strength and durability of the knobs, and should prevent this type of event from recurring. No further action is required.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1526439-2001-00060
MDR Report Key364193
Report Source08
Date Received2001-12-05
Date of Report2001-12-05
Date Mfgr Received2001-11-05
Date Added to Maude2001-12-07
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location0
Manufacturer Street325 PARAMOUNT DRIVE
Manufacturer CityRAYNHAM MA 02767
Manufacturer CountryUS
Manufacturer Postal02767
Manufacturer Phone5088808100
Manufacturer G1*
Manufacturer Street*
Manufacturer City*
Manufacturer Country*
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameBREMER HALO SYSTEM
Generic NamePEDIATRIC HALO
Product CodeKQZ
Date Received2001-12-05
Model NumberNA
Catalog NumberPT002
Lot Number043136
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Implant FlagN
Date RemovedU
Device Sequence No1
Device Event Key353294
ManufacturerDEPUY ACROMED, INC.
Manufacturer Address325 PARAMOUNT DR. RAYNHAM MA 02767 US
Baseline Brand NameBREMER HALO SYSTEM
Baseline Generic NamePEDIATRIC HALO
Baseline Model NoNA
Baseline Catalog NoPT002
Baseline IDNA


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2001-12-05

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.