VITROS CHEMISTRY PRODUCTS LI SLIDES 1632660

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 00,01,05 report with the FDA on 2014-02-28 for VITROS CHEMISTRY PRODUCTS LI SLIDES 1632660 manufactured by Ortho-clinical Diagnostics.

Event Text Entries

[19021172] A lower than expected vitros li result (qc level ii < 0. 20 versus expected 1. 09 mmol/l) was obtained from a quality control fluid processed on a vitros 250 chemistry system. Biased results of the magnitude and direction observed may lead to inappropriate physician action if patient samples were affected. There was no report of patient harm as a result of this event. This report corresponds to ortho clinical diagnostics inc. Complaint number (b)(4).
Patient Sequence No: 1, Text Type: D, B5


[19128591] The investigation determined that a lower than expected vitros li result was obtained from a qc fluid run on a vitros 250 chemistry system. The evaluation of historical laboratory results demonstrated that reagent lot in question had good accuracy prior to the date of the event. The investigation determined that the issue was isolated to a specific vitros li slide cartridge that was suspected to be non-reactive. Acceptable vitros li performance was achieved using an alternate slide cartridge without any changes being made to the vitros 250 chemistry system. The assignable cause of the non-reactive slide cartridge could not be determined.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1319809-2014-00011
MDR Report Key3653652
Report Source00,01,05
Date Received2014-02-28
Date of Report2014-02-28
Date of Event2013-12-25
Date Mfgr Received2014-02-27
Device Manufacturer Date2012-08-08
Date Added to Maude2014-05-12
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMR. JAMES A STEVENS
Manufacturer Street100 INDIGO CREEK DRIVE
Manufacturer CityROCHESTER NY 14626
Manufacturer CountryUS
Manufacturer Postal14626
Manufacturer Phone5854533000
Manufacturer G1ORTHO-CLINICAL DIAGNOSTICS
Manufacturer Street513 TECHNOLOGY BLVD.
Manufacturer CityROCHESTER NY 14652
Manufacturer CountryUS
Manufacturer Postal Code14652
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameVITROS CHEMISTRY PRODUCTS LI SLIDES
Generic NameIN-VITRO DOAGNOSTIC
Product CodeJIH
Date Received2014-02-28
Catalog Number1632660
Lot Number3119-0091-1725
Device Expiration Date2014-02-01
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerORTHO-CLINICAL DIAGNOSTICS
Manufacturer Address513 TECHNOLOGY BLVD. ROCHESTER NY 14652 US 14652


Patients

Patient NumberTreatmentOutcomeDate
10 2014-02-28

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.