PKG; 26IN VISION ELECT HDTV SURGICAL VIEWING MONITOR 0240030960

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 04 report with the FDA on 2014-03-03 for PKG; 26IN VISION ELECT HDTV SURGICAL VIEWING MONITOR 0240030960 manufactured by Stryker Endoscopy-san Jose.

Event Text Entries

[4450019] It was reported that there was a shock.
Patient Sequence No: 1, Text Type: D, B5


[11788389] The product was not returned for investigation. The reported failure mode could not be confirmed. Probable root causes for the reported failure could have been caused by the power supply, ac inlet fuse, and/or insulation. However, this cannot be confirmed since the unit was not returned. In the event that the unit is returned, a full evaluation will be conducted and a follow up report will be issued. In sum, the product was not returned for investigation and the reported failure mode could not be confirmed. The failure mode will be monitored for future reoccurrence.
Patient Sequence No: 1, Text Type: N, H10


[17927989] It was reported that there was a shock.
Patient Sequence No: 1, Text Type: D, B5


[18152523] Additional information will be provided once the investigation has been completed.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number0002936485-2014-00101
MDR Report Key3656154
Report Source04
Date Received2014-03-03
Date of Report2014-02-05
Date of Event2014-01-15
Date Mfgr Received2014-02-05
Date Added to Maude2014-05-13
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMR. THOMAS SHAFER
Manufacturer Street5900 OPTICAL COURT
Manufacturer CitySAN JOSE CA 95138
Manufacturer CountryUS
Manufacturer Postal95138
Manufacturer Phone4087542000
Manufacturer G1STRYKER ENDOSCOPY-SAN JOSE
Manufacturer Street5900 OPTICAL COURT
Manufacturer CitySAN JOSE CA 95138
Manufacturer CountryUS
Manufacturer Postal Code95138
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NamePKG; 26IN VISION ELECT HDTV SURGICAL VIEWING MONITOR
Generic NameCAMERA, TELEVISION, SURGICAL, WITHOUT AUDIO
Product CodeFWB
Date Received2014-03-03
Catalog Number0240030960
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerSTRYKER ENDOSCOPY-SAN JOSE
Manufacturer Address5900 OPTICAL COURT SAN JOSE CA 95138 US 95138


Patients

Patient NumberTreatmentOutcomeDate
10 2014-03-03

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.