ZEON APOLLO ZZ-APOLLO-TTL

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2014-03-07 for ZEON APOLLO ZZ-APOLLO-TTL manufactured by Kerr Corporation.

Event Text Entries

[4321202] A complainant alleged that the zeon apollo device got hot, slightly burning the desk, and the light bezzle came off.
Patient Sequence No: 1, Text Type: D, B5


[11689001] No injuries were associated with this incident; the complainant is doing fine. To date, the product has not been returned; therefore, no evaluation can be conducted. This incident is mdr reportable as a malfunction because the device could cause or contribute to a serious injury if the malfunction were to recur.
Patient Sequence No: 1, Text Type: N, H10


[31117666] The unit was returned and a visual and physical inspection revealed that there were no damages to the battery pack and the charger. Contrary to the alleged complaint, both items were working properly. It was revealed that the headlight was exposed to an excessive amount of heat causing the headlight to melt. After reproducing the failure with the apollo components, it is concluded that the apollo battery pack is not capable of producing a sufficient amount of current that caused the plastic lens in the headlight to melt. A power source capable of producing about five (5) times more power than the apollo pack is required in order to melt the plastic lens of the headlight.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3003848022-2014-00002
MDR Report Key3667409
Report Source05
Date Received2014-03-07
Date of Report2014-02-11
Date of Event2014-02-11
Date Mfgr Received2014-02-11
Date Added to Maude2014-05-14
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationDENTAL ASSISTANT
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMRS. KERRI CASINO
Manufacturer Street1717 W. COLLINS AVE
Manufacturer CityORANGE CA 92867
Manufacturer CountryUS
Manufacturer Postal92867
Manufacturer Phone7145167634
Manufacturer G1KERR CORPORATION
Manufacturer Street3225 DEMING WAY SUITE 190
Manufacturer CityMIDDLETON WI 53562
Manufacturer CountryUS
Manufacturer Postal Code53562
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameZEON APOLLO
Generic NameSURGICAL HEADLIGHT
Product CodeEBA
Date Received2014-03-07
Catalog NumberZZ-APOLLO-TTL
OperatorDENTAL ASSISTANT
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerKERR CORPORATION
Manufacturer Address3225 DEMING WAY SUITE 190 MIDDLETON WI 53562 US 53562


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2014-03-07