ONLINE DAT BENZODIAZEPINES PLUS 04490789190

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2014-03-10 for ONLINE DAT BENZODIAZEPINES PLUS 04490789190 manufactured by Roche Diagnostics.

Event Text Entries

[16913508] The customer received questionable results for benzodiazepines plus (benz) on one patient sample. The initial benz result was negative using the qualitative 200 ng/ml cutoff screening test. The result was reported outside of the laboratory. The sample was sent to a reference laboratory and was tested by gas chromatography/mass spectrometry (gc/ms) and generated a result of 1557 ng/ml lorazepam in the sample. The customer deemed the gc/ms result to be the correct result. There was no adverse event. The lot number for the benz reagent in use was 68216501, with an expiration date of 01/31/2015. The field service representative found the gear pump pressure was low. He adjusted the pressure to specification. He performed mechanism checks, which passed. The customer performed calibration and qc successfully.
Patient Sequence No: 1, Text Type: D, B5


[16966691] It was unknown if the initial reporter sent report to the fda.
Patient Sequence No: 1, Text Type: N, H10


[18763948] The investigation determined that lorazepam metabolites have a 0. 5% cross reactivity with the benz assay. Since lorazepam metabolites are frequently present in the urine, the assay will not detect it with the found concentration. This is indicated in the product labeling.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1823260-2014-01673
MDR Report Key3670532
Report Source05,06
Date Received2014-03-10
Date of Report2014-03-21
Date of Event2014-02-19
Date Mfgr Received2014-02-27
Date Added to Maude2014-03-10
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1ROCHE DIAGNOSTICS GMBH
Manufacturer StreetSANDHOFERSTRASSE 116 NA
Manufacturer CityMANNHEIM (BADEN-WURTTEMBERG) 68305
Manufacturer CountryGM
Manufacturer Postal Code68305
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameONLINE DAT BENZODIAZEPINES PLUS
Generic NameENZYME IMMUNOASSAY, BENZODIAZIPINE
Product CodeJXM
Date Received2014-03-10
Model NumberNA
Catalog Number04490789190
Lot Number68216501
ID NumberNA
Device Expiration Date2015-01-31
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 46250 US 46250


Patients

Patient NumberTreatmentOutcomeDate
10 2014-03-10

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.