VALEO OL/PL SPINE FUSION 91.004.3710

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2014-03-10 for VALEO OL/PL SPINE FUSION 91.004.3710 manufactured by Amedica.

Event Text Entries

[4259773] A medica box chisel sets for valeo ol/pl spine fusion designed with small interior ridges that make adequate clearing very difficult and questionable. Item found with bone debris intraoperatively. Took multiple attempts, under high magnification to adequately clean bone fragments. Design reflected in all chisels in this series.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report NumberMW5034917
MDR Report Key3677447
Date Received2014-03-10
Date of Report2014-03-07
Date of Event2014-03-05
Date Added to Maude2014-03-14
Event Key0
Report Source CodeVoluntary report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Details

Brand NameVALEO OL/PL SPINE FUSION
Generic NameBONE CHISEL
Product CodeEML
Date Received2014-03-10
Model Number91.004.3710
OperatorHEALTH PROFESSIONAL
Device Sequence No1
Device Event Key0
ManufacturerAMEDICA
Manufacturer AddressSALT LAKE CITY UT 84119 US 84119


Patients

Patient NumberTreatmentOutcomeDate
10 2014-03-10

© 2026 FDA.report