MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2014-03-20 for DIMENSION VISTA CLINICAL CHEMISTRY SYSTEM DIMENSION VISTA 1500 manufactured by Siemens Healthcare Diagnostics Inc..
[4258001]
A discordant, falsely elevated potassium result was obtained on one patient sample on a dimension vista 1500 instrument. The sample was flagged and automatically repeated and resulted lower than the initial result, but the value was still elevated. The discordant result(s) was reported to the physician(s), who treated the patient. The sample was repeated on the same instrument and resulted lower than both of the previous results. There are no known reports of adverse health consequences due to the discordant, falsely elevated potassium results.
Patient Sequence No: 1, Text Type: D, B5
[11590814]
A siemens global product support specialist evaluated, the instrument data and did not find an instrument malfunction. The cause of the discordant, falsely elevated potassium results is unknown. It was discovered that the customer centrifuges patient samples outside of the tube vendor specifications, which is a failure to follow instructions. The instrument is performing according to specifications. No further evaluation of this device is required. Medwatch # (b)(4) was filed by the customer.
Patient Sequence No: 1, Text Type: N, H10
Report Number | 1226181-2014-00153 |
MDR Report Key | 3691330 |
Report Source | 05,06 |
Date Received | 2014-03-20 |
Date of Report | 2014-03-04 |
Date of Event | 2014-01-12 |
Date Mfgr Received | 2014-03-04 |
Device Manufacturer Date | 2011-06-13 |
Date Added to Maude | 2014-05-20 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 0 |
Event Location | 0 |
Manufacturer Contact | CASSANDRA KOCSIS |
Manufacturer Street | 511 BENEDICT AVE |
Manufacturer City | TARRYTOWN NY 10591 |
Manufacturer Country | US |
Manufacturer Postal | 10591 |
Manufacturer Phone | 9145242687 |
Manufacturer G1 | SIEMENS HEALTHCARE DIAGNOSTICS INC. |
Manufacturer Street | 101 SILVERMINE ROAD |
Manufacturer City | BROOKFIELD CT 06804 |
Manufacturer Country | US |
Manufacturer Postal Code | 06804 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | DIMENSION VISTA CLINICAL CHEMISTRY SYSTEM |
Generic Name | CLINICAL CHEMISTRY ANALYZER |
Product Code | MZV |
Date Received | 2014-03-20 |
Model Number | DIMENSION VISTA 1500 |
Operator | HEALTH PROFESSIONAL |
Device Availability | N |
Device Eval'ed by Mfgr | N |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | SIEMENS HEALTHCARE DIAGNOSTICS INC. |
Manufacturer Address | 101 SILVERMINE RD BROOKFIELD CT 06804 US 06804 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2014-03-20 |