HEARTMATE RECHARGEABLE BATTERY 102515

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2013-12-17 for HEARTMATE RECHARGEABLE BATTERY 102515 manufactured by Thoratec Corp..

Event Text Entries

[4259608] The patient was implanted with a left ventricular assist device (lvad). It was reported by the transplant coordinator that multiple power cable disconnect and red heart alarms were noted in patient history. A sign of corrosion on one of the patient's batteries was noted. The hospital replaced the patient's batteries and battery clips. Additional information was received 6 days later indicating that with the change of batteries and battery clips, no alarms were occurring.
Patient Sequence No: 1, Text Type: D, B5


[11604980] The patient remains on going with the implanted device and no further issues have been reported. No further information is available at this time. The manufacturer was advised that the product will not be returned for evaluation. No further information is available at this time. A supplemental report will be submitted when the manufacturer's investigation is completed.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2916596-2013-01667
MDR Report Key3694589
Report Source05,06
Date Received2013-12-17
Date of Report2013-11-19
Date of Event2013-11-19
Date Facility Aware2013-11-19
Report Date2013-11-19
Date Reported to Mfgr2013-11-19
Date Mfgr Received2013-11-19
Date Added to Maude2014-04-15
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location3
Manufacturer ContactMR ROBERT FRYC
Manufacturer Street23 FOURTH AVE.
Manufacturer CityBURLINGTON MA 01803
Manufacturer CountryUS
Manufacturer Postal01803
Manufacturer Phone7812720139
Manufacturer G1THORATEC CORP.
Manufacturer Street6035 STONERIDGE DR.
Manufacturer CityPLEASANTON CA 94588
Manufacturer CountryUS
Manufacturer Postal Code94588
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameHEARTMATE RECHARGEABLE BATTERY
Generic NameRECHARGEABLE BATTERY
Product CodeMOY
Date Received2013-12-17
Model Number102515
Catalog Number102515
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerTHORATEC CORP.
Manufacturer AddressPLEASANTON CA US


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2013-12-17