AU680 CLINICAL CHEMISTRY ANALYZER AU681-10E B12188

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2014-03-25 for AU680 CLINICAL CHEMISTRY ANALYZER AU681-10E B12188 manufactured by Beckman Coulter.

Event Text Entries

[4246569] A customer stated they generated 2 erroneously high potassium results on their au680 analyser. These results were not reported outside of the laboratory so there was no reported change to patient treatment. According to the customer, quality control runs were within specification and all required maintenance had been performed on the ise unit.
Patient Sequence No: 1, Text Type: D, B5


[11594192] A field service engineer (fse) was dispatched. The fse found large amount of bubbles coming from the reference valve. The fse replaced the reference valve with part no. B37620. The fse primed the system and confirmed there were no further bubbles coming from the valve. A calibration and qc were performed and the system was confirmed to be performing with specification. (b)(4).
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number9612296-2014-00052
MDR Report Key3699621
Report Source05,06
Date Received2014-03-25
Date of Report2014-03-08
Date of Event2014-03-08
Date Mfgr Received2014-03-08
Device Manufacturer Date2012-03-09
Date Added to Maude2014-05-22
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMS. DUNG NGUYEN
Manufacturer Street250 S. KRAEMER BLVD.
Manufacturer CityBREA CA 92821
Manufacturer CountryUS
Manufacturer Postal92821
Manufacturer Phone7149614941
Manufacturer G1BECKMAN COULTER MISHIMA K.K.
Manufacturer Street454-32 HIGASHINO NAGAIZUMI-CHO
Manufacturer CitySUNTO-GUN, SCHIZUOKA 411-093
Manufacturer CountryJA
Manufacturer Postal Code411-093
Single Use3
Remedial ActionMA
Previous Use Code3
Removal Correction Number2050012-08/22/2013-008
Event Type3
Type of Report3

Device Details

Brand NameAU680 CLINICAL CHEMISTRY ANALYZER
Generic NameANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE
Product CodeMZV
Date Received2014-03-25
Model NumberAU681-10E
Catalog NumberB12188
Lot NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBECKMAN COULTER
Manufacturer Address250 S. KRAEMER BLVD. BREA CA 92821 US 92821


Patients

Patient NumberTreatmentOutcomeDate
10 2014-03-25

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.