Event Text Entries

[20974271] Burn.? 'a little boy was recently burned when using several different tebbetts retractors. The customer says that the light source connection has been hot at the light source and burning the drapes for about a month. They have been using a light cord that has cracks in the insulation. Customer was told to stop using any light cord that has cracks in the insulation @ that they need to use a compatible light cord which (b)(6) verified. They have been using 88-9727 - light cord 5mm, light source : luxtec, they use tebbetts retractors 88-1187, 88-1088 moving forward, they will use compatible light cord 88-9760 - 3. 5mm cable': additional information received from the sales rep (b)(4) 2014. It was reported that the customer stated that the tebbetts retractors are ok and they are just getting hot cause they were using a codd that was delivering too much energy. We will not be receiving the instruments for evaluation.
Patient Sequence No: 1, Text Type: D, B5


[21280160] (b)(4) should additional information be received, a follow-up emdr will be submitted.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1038548-2014-00037
MDR Report Key3701353
Report Source04
Date Received2014-03-26
Date of Report2014-03-26
Date of Event2014-03-04
Date Mfgr Received2014-03-04
Date Added to Maude2014-03-26
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactJILL RITTORNO
Manufacturer Street75 NORTH FAIRWAY DRIVE
Manufacturer CityVERNON HILLS IL 60061
Manufacturer CountryUS
Manufacturer Postal60061
Manufacturer Phone8473628056
Manufacturer G1CAREFUSION 2200, INC. SNOWDEN PENCER PRODUCTS
Manufacturer Street5175 SOUTH ROYAL ATLANTA DR
Manufacturer CityTUCKER 30084
Manufacturer CountryUS
Manufacturer Postal Code30084
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameUNK V. MUELLER
Generic NameRETRACTOR, FIBEROPTIC
Product CodeFDG
Date Received2014-03-26
Model NumberUNK V. MUELLER
OperatorPHYSICIAN
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerCAREFUSION
Manufacturer Address75 NORTH FAIRWAY DRIVE VERNON HILLS IL 60061 US 60061


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2014-03-26