SILICONE ELASTOMER COATED LATEX CATHETER TRAY DYND160116

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2014-01-13 for SILICONE ELASTOMER COATED LATEX CATHETER TRAY DYND160116 manufactured by Medline Industries, Inc..

Event Text Entries

[20781613] The catheter balloon would not inflate. A new catheter was then inserted without further incident.
Patient Sequence No: 1, Text Type: D, B5


[21001973] Prior to the insertion of the foley catheter, the clinician tested the balloon by inflating it with 10 ml of saline. It apparently inflated and deflated without issue. She then inserted the catheter and attempted to inflate the balloon. After inserting 5 ml of saline, she noticed a bulging at the y of the tubing. She removed the catheter and then inserted a new catheter without further incident. The sample was returned and evaluated. The inflation lumen was found to be collapsed which prevented the inflation of the balloon. The root cause for the collapse was not determined. We pulled samples from stock and tested them. We were unable to replicate the reported issue. The balloons were tested and all inflated and deflated according to specification.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1417592-2014-00001
MDR Report Key3708940
Report Source06
Date Received2014-01-13
Date of Report2014-01-03
Date of Event2013-12-05
Date Mfgr Received2013-12-05
Device Manufacturer Date2013-08-01
Date Added to Maude2014-03-31
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationRISK MANAGER
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location3
Manufacturer ContactJULIE FINLEY
Manufacturer StreetONE MEDLINE PLACE
Manufacturer CityMUNDELEIN IL 60060
Manufacturer CountryUS
Manufacturer Postal60060
Manufacturer Phone8476434709
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSILICONE ELASTOMER COATED LATEX CATHETER TRAY
Product CodeNWR
Date Received2014-01-13
Catalog NumberDYND160116
Lot Number13HB4297
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerMEDLINE INDUSTRIES, INC.
Manufacturer AddressMUNDELEIN IL US


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2014-01-13

© 2026 FDA.report