INJECTION NEEDLE MCL55-3 240MM BAYONET [MXI/XOM]

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,06,07 report with the FDA on 2014-03-28 for INJECTION NEEDLE MCL55-3 240MM BAYONET [MXI/XOM] manufactured by Xomed Microfrance Mfg.

Event Text Entries

[17158199] It was reported that a 240mm bayonet injection needle broke. However, analysis of the complaint device confirmed the break resulted in a 3cm device fragment that was not returned by the customer. There was no report of patient impact or injury as a result of this event.
Patient Sequence No: 1, Text Type: D, B5


[17365883] (b)(4). In response to medtronic? S request for device return, the device was returned to medtronic? S international manufacturing site for evaluation purposes. The investigation report findings are detailed as follows: the needle is broken right after the laser weld? The second part was not returned. No manufacturing or material defect was observed. The most probable cause of this breakage is an excessive effort during use or the reprocessing of the instrument. It is very fragile and in the case of a deformation, it is impossible to straighten it without breaking the tube. This device is used for therapeutic purposes.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number9680837-2014-00021
MDR Report Key3709371
Report Source01,06,07
Date Received2014-03-28
Date of Report2014-03-03
Date of Event2014-02-26
Date Mfgr Received2014-03-03
Device Manufacturer Date2013-10-01
Date Added to Maude2014-05-28
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactJACKLYN HAYMAN
Manufacturer Street6743 SOUTHPOINT DRIVE NORTH
Manufacturer CityJACKSONVILLE FL 32216
Manufacturer CountryUS
Manufacturer Postal32216
Manufacturer Phone9042812769
Manufacturer G1MEDTRONIC XOMED, INC.
Manufacturer Street6743 SOUTHPOINT DRIVE NORTH
Manufacturer CityJACKSONVILLE FL 32216
Manufacturer CountryUS
Manufacturer Postal Code32216
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameINJECTION NEEDLE MCL55-3 240MM BAYONET [MXI/XOM]
Generic NameNEEDLE, ASPIRATION AND INJECTION, REUSABLE
Product CodeGDM
Date Received2014-03-28
Returned To Mfg2014-03-03
Model NumberMCL55-3
Catalog NumberMCL55-3
Lot Number131001
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerXOMED MICROFRANCE MFG
Manufacturer AddressSAINT-AUBIN-LE-MONIAL BOURBON-L'ARCHAMBAULT 3160 FR 3160


Patients

Patient NumberTreatmentOutcomeDate
10 2014-03-28

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.