XTRATRACE BELT 1421040

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2014-02-26 for XTRATRACE BELT 1421040 manufactured by Embla Systems.

Event Text Entries

[4330118] It was observed that a pt has red welt like mark where the xacttrace belts were on his back only. Pt went to emergency.
Patient Sequence No: 1, Text Type: D, B5


[11568187] In compliance with our quality management system requested the client to send back the xactrabelt for evaluation. Functional and temperature tests were performed and all results indicate that the device is working within the specifications - refer to the enclosed tests report. Preliminary review of the event details revealed that pt was not wearing a shirt while using the xactrabelt. The product labeling (attached herewith) clearly instructs that the exactrabelt is intended to be worn over the nightclothes. Based on the evaluation resulted and the event information we concluded that this incident was caused by the user not following the manufacturer's instructions. The client is now informed and advised to ensure that product labeling instructions are followed.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3005888786-2014-00001
MDR Report Key3723906
Report Source06
Date Received2014-02-26
Date of Report2014-03-25
Date of Event2014-02-28
Date Mfgr Received2014-02-28
Device Manufacturer Date2013-10-01
Date Added to Maude2014-04-04
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactSANJAY MEHTA
Manufacturer Street1 HINES ROAD
Manufacturer CityKANATA K2K3C7
Manufacturer CountryCA
Manufacturer PostalK2K3C7
Manufacturer Phone2875055
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameXTRATRACE BELT
Generic NameUNIVERSAL XACTTRACE
Product CodeMNR
Date Received2014-02-26
Returned To Mfg2014-03-11
Model Number1421040
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerEMBLA SYSTEMS
Manufacturer Address1 HINES ROAD KANATA K2K3C7 CA K2K3C7


Patients

Patient NumberTreatmentOutcomeDate
10 2014-02-26