MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,04,07 report with the FDA on 2014-02-27 for APPLIED BIOSYSTEMS 7500 FAST X REAL-TIME PCR 4407205 manufactured by Life Technologies Holdings Pte. Ltd..
[4325395]
A customer reported an overheated plastic smell coming from the instrument, applied biosystems 7500 fast dx (cat. No. # 4407205, serial no. (b)(4)). No pt involvement reported.
Patient Sequence No: 1, Text Type: D, B5
[11595299]
Device is an instrument and is not implanted/ explanted. The instrument was investigated on site. The complained issue could not be reproduced. While checking the instrument components, a slight discoloration on the lamp cable to the lamp holder connector was observed. It was repaired on site by replacing the lamp holder, lamp cable and lamp. Oq/pv was performed after the repair and the instrument passed required specifications. Device intended use: the applied biosystems 7500 fast dx real-time pcr instrument with the sds software version 1. 4 is a real-time nucleic acid amplification and detection system that measures nucleic acid signals from reverse transcribed rna and converts them to comparative quantitative readouts using fluorescent detection of dual-labeled hydrolysis probes. The 7500 fast dx real-time pcr instrument is to be used only by technologists trained in laboratory techniques, procedures, and on use of the analyzer.
Patient Sequence No: 1, Text Type: N, H10
Report Number | 3003673482-2014-00002 |
MDR Report Key | 3727142 |
Report Source | 01,04,07 |
Date Received | 2014-02-27 |
Date of Report | 2014-02-07 |
Date of Event | 2014-02-07 |
Date Mfgr Received | 2014-02-07 |
Device Manufacturer Date | 2009-11-01 |
Date Added to Maude | 2014-05-05 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 0 |
Event Location | 0 |
Manufacturer Contact | KELLI TANZELLA |
Manufacturer Street | 3175 STALEY RD. |
Manufacturer City | GRAND ISLAND NY 14072 |
Manufacturer Country | US |
Manufacturer Postal | 14072 |
Manufacturer Phone | 7167743122 |
Manufacturer G1 | LIFE TECHNOLOGIES HOLDINGS PTE. LTD. |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | APPLIED BIOSYSTEMS 7500 FAST X REAL-TIME PCR |
Product Code | NSU |
Date Received | 2014-02-27 |
Catalog Number | 4407205 |
Lot Number | 275030204 |
Operator | HEALTH PROFESSIONAL |
Device Availability | N |
Device Age | DA |
Device Eval'ed by Mfgr | R |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | LIFE TECHNOLOGIES HOLDINGS PTE. LTD. |
Manufacturer Address | SINGAPORE SN |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2014-02-27 |