CAPTURE-R SELECT 0006446

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2014-04-08 for CAPTURE-R SELECT 0006446 manufactured by Immucor, Inc..

Event Text Entries

[20696708] On (b)(6) 2014, a customer reported unexpected compatible crossmatch results when using capture-r select on a galileo neo instrument on (b)(6) 2014.
Patient Sequence No: 1, Text Type: D, B5


[21002372] Immucor technical support used remote electronic access method to assess the galileo neo electronic images on (b)(4) 2014. The well images for the unexpected crossmatch compatible wells all appeared as unexpected negative with associated tightly defined red blood cell buttons. Immucor laboratory tested retention products on (b)(4) 2014 and all retention products performed as expected. The customer returned blood samples to immucor for testing, and on (b)(4) 2014 the immucor laboratory duplicated the unexpected compatible crossmatch results. The unexpected crossmatch compatible results have a potential for a transfusion error.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1034569-2014-00057
MDR Report Key3734631
Report Source05,06
Date Received2014-04-08
Date of Report2014-04-07
Date of Event2014-03-11
Date Mfgr Received2014-03-14
Device Manufacturer Date2014-01-25
Date Added to Maude2014-06-03
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationMEDICAL TECHNOLOGIST
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMR HOWARD YOREK
Manufacturer Street3130 GATEWAY DRIVE
Manufacturer CityNORCROSS GA 300915625
Manufacturer CountryUS
Manufacturer Postal300915625
Manufacturer Phone7704412051
Manufacturer G1IMMUCOR, INC.
Manufacturer Street3130 GATEWAY DRIVE
Manufacturer CityNORCROSS GA 30091562
Manufacturer CountryUS
Manufacturer Postal Code30091 5625
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCAPTURE-R SELECT
Generic NameSOLID PHASE SYSTEM FOR THE IMMOBILIZATION OF HUMAN ERYTHROCYTES
Product CodeKSS
Date Received2014-04-08
Catalog Number0006446
Lot NumberSC262
Device Expiration Date2015-01-25
OperatorMEDICAL TECHNOLOGIST
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerIMMUCOR, INC.
Manufacturer Address3130 GATEWAY DRIVE NORCROSS GA 30091562 US 30091 5625


Patients

Patient NumberTreatmentOutcomeDate
10 2014-04-08

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.