OLYMPUS DISPOSABLE BIOPSY FORCEPS FB-221D

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2014-01-22 for OLYMPUS DISPOSABLE BIOPSY FORCEPS FB-221D manufactured by Olympus Medical System Corp..

Event Text Entries

[4551291] Olympus was informed that during four to six therapeutic bronchoscopy procedures, plastic shavings were noted in the patient's lungs. The shavings were thin and approximately 5mm. The shavings were retrieved with forceps or suction. The user did feel some resistance when inserting the forceps into the biopsy channel. There was no patient injury reported. Olympus followed up with the user facility to obtain additional information regarding the report and was informed that the shavings were attributed to the needle of the forceps not retracting when the forceps were closed. The exposed needle shaved the inside of the biopsy channel during deployment.
Patient Sequence No: 1, Text Type: D, B5


[11833546] An olympus representative visited the user facility to inspect the forceps and noted that the needle was bent outward. The forceps will be returned to olympus for further investigation.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2951238-2014-00018
MDR Report Key3747820
Report Source06
Date Received2014-01-22
Date of Report2014-01-06
Date of Event2013-12-27
Date Mfgr Received2014-01-06
Date Added to Maude2014-05-01
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactLAURA STORMS
Manufacturer Street2400 RINGWOOD AVE.
Manufacturer CitySAN JOSE CA 95131
Manufacturer CountryUS
Manufacturer Postal95131
Manufacturer Phone4848965688
Manufacturer G1OLYMPUS MEDICAL SYSTEM CORP.
Manufacturer Street2951 ISHIKAWA-CHO
Manufacturer CityHACHIOJI-SHI, TOKYO 192-8507
Manufacturer CountryJA
Manufacturer Postal Code192-8507
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameOLYMPUS DISPOSABLE BIOPSY FORCEPS
Generic NameDISPOSABLE BIOPSY FORCEPS
Product CodeHFB
Date Received2014-01-22
Model NumberFB-221D
Catalog NumberFB-221D
Lot Number38T
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerOLYMPUS MEDICAL SYSTEM CORP.
Manufacturer Address2951 ISHIKAWA-CHO HACHIOJI-SHI, TOKYO 192-8507 JA 192-8507


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2014-01-22

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.