EASYSTAND STRAPSTAND 2000

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2014-01-24 for EASYSTAND STRAPSTAND 2000 manufactured by Altimate Medical, Inc..

Event Text Entries

[4547143] Received a call from a facility that the stitching came loose on the large strap, part no. 80719 on their easystand strapstand standing frame. The facility has had the unit since (b)(6) 2007 and they use the unit every day with approximately 10 different residents. The user's wheelchair is positioned up to the standing frame so it's available when the user goes back to the seated position, therefore, no injuries were reported when the stitching came loose.
Patient Sequence No: 1, Text Type: D, B5


[11832521] After discussing this issue further with the facility, it was determined that this facility has had the standing frame for approximately 7 years and uses the device on a daily basis with 10 different residents. It is assumed that the stitching came loose on this strap due to amount of time that this product has been used. The facility stated that overall the product is in good condition and that they would like two replacement straps due to the amount that this product is used. Altimate provided the facility with two replacement straps per their request.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2183634-2014-00001
MDR Report Key3748051
Report Source06
Date Received2014-01-24
Date of Report2014-01-07
Date of Event2014-01-07
Date Mfgr Received2014-01-07
Device Manufacturer Date2007-02-01
Date Added to Maude2014-05-08
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactSTACEY FRANK
Manufacturer Street262 WEST FIRST ST.
Manufacturer CityMORTON MN 56270
Manufacturer CountryUS
Manufacturer Postal56270
Manufacturer Phone5076976393
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameEASYSTAND STRAPSTAND
Generic Name890.5370
Product CodeION
Date Received2014-01-24
Returned To Mfg2014-01-16
Model NumberSTRAPSTAND
Catalog Number2000
Lot NumberNA
OperatorOTHER
Device AvailabilityY
Device AgeDA
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerALTIMATE MEDICAL, INC.
Manufacturer Address262 WEST FIRST ST. MORTON MN 56270 US 56270


Patients

Patient NumberTreatmentOutcomeDate
10 2014-01-24

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.