MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2013-11-22 for ICHEM 100 800-7005 manufactured by Iris International.
[4465250]
False positive leukocyte esterase was reported on 101 patient results which lead to 12 patients to receive unnecessary antibiotics. The malfunction was confirmed by urine microscopy and urine cultures which were negative for urinary tract infection in all 12 patients.
Patient Sequence No: 1, Text Type: D, B5
[11795465]
The following mdrs are related to this event: 2023446-2013-00001, 2023446-2013-00003, 2023446-2013-00004, 2023446-2013-00005, 2023446-2013-00006, 2023446-2013-00007, 2023446-2013-00008, 2023446-2013-00009, 2023446-2013-00010, 2023446-2013-00011, 2023446-2013-00012, 2023446-2013-00013. See scanned page.
Patient Sequence No: 1, Text Type: N, H10
Report Number | 2023446-2013-00013 |
MDR Report Key | 3748670 |
Report Source | 05 |
Date Received | 2013-11-22 |
Date of Report | 2013-11-22 |
Date of Event | 2013-08-15 |
Date Mfgr Received | 2013-08-15 |
Device Manufacturer Date | 2013-04-01 |
Date Added to Maude | 2014-05-08 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 0 |
Reporter Occupation | OTHER HEALTH CARE PROFESSIONAL |
Health Professional | 0 |
Initial Report to FDA | 3 |
Report to FDA | 0 |
Event Location | 0 |
Manufacturer Contact | SUDHA GUPTA |
Manufacturer Street | 9172 ETON AVE. |
Manufacturer City | CHATSWORTH CA 91311 |
Manufacturer Country | US |
Manufacturer Postal | 91311 |
Manufacturer Phone | 8185277497 |
Single Use | 0 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | ICHEM 100 |
Generic Name | URINE CHEMISTRY STRIPS |
Product Code | JJW |
Date Received | 2013-11-22 |
Model Number | 800-7005 |
Lot Number | 510034A |
Device Expiration Date | 2015-04-01 |
Operator | HEALTH PROFESSIONAL |
Device Eval'ed by Mfgr | R |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | IRIS INTERNATIONAL |
Manufacturer Address | CHATSWORTH CA 91311 US 91311 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 1. Other | 2013-11-22 |