UNICEL? DXC 600 SYNCHRON? SYSTEM A10405

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2014-04-14 for UNICEL? DXC 600 SYNCHRON? SYSTEM A10405 manufactured by Beckman Coulter.

Event Text Entries

[4486815] A customer reported their dxc 600 instrument generated false high k (potassium) and cl (chloride) results for 3 patient samples. Na (sodium), co2 (carbon dioxide) and calc (calcium) results were suppressed (no results generated) for these same samples. The results were not reported out of the laboratory so there was no impact to patient treatment. When the issue was noticed, samples were repeated. It is unknown if the repeats were performed on the same or another instrument.
Patient Sequence No: 1, Text Type: D, B5


[11841760] A field service engineer was not dispatched. The customer support representative assisted the customer over the phone. The customer found the end of line #24 (ise buffer reagent) was not connected at the ratio pump. The customer trimmed and reconnected the line, resolving the issue. (b)(4). Customer was able to resolve the issue.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2050012-2014-00200
MDR Report Key3749752
Report Source05,06
Date Received2014-04-14
Date of Report2014-03-27
Date of Event2014-03-27
Date Mfgr Received2014-03-27
Device Manufacturer Date2005-09-24
Date Added to Maude2014-06-06
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationMEDICAL TECHNOLOGIST
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMS. DUNG NGUYEN
Manufacturer Street250 S. KRAEMER BLVD.
Manufacturer CityBREA CA 92821
Manufacturer CountryUS
Manufacturer Postal92821
Manufacturer Phone7149614941
Manufacturer G1BECKMAN COULTER
Manufacturer Street250 S. KRAEMER BLVD.
Manufacturer CityBREA CA 92821
Manufacturer CountryUS
Manufacturer Postal Code92821
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameUNICEL? DXC 600 SYNCHRON? SYSTEM
Generic NameANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE
Product CodeMZV
Date Received2014-04-14
Model NumberNA
Catalog NumberA10405
Lot NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerBECKMAN COULTER
Manufacturer Address250 S. KRAEMER BLVD. BREA CA 92821 US 92821


Patients

Patient NumberTreatmentOutcomeDate
10 2014-04-14

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.