SENSE BREAST 3.0T 4535-302-64711

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 08 report with the FDA on 2014-01-31 for SENSE BREAST 3.0T 4535-302-64711 manufactured by Invivo Corporation.

Event Text Entries

[4464241] The customer reported that during a test of an mri breast protocol with the new mammocoil on the philips pet/mri, a patient was positioned on the coil and the table was moved into the mri tunnel using the "travel to scanplane" button. During the movement of the table to the scanning position, the patient became trapped between the coil and the top of the bore. The patient indicated pain and was immediately removed from the bore. A few days later the patient was diagnosed with a broken rib.
Patient Sequence No: 1, Text Type: D, B5


[11737613] (b)(4): eval method: the investigation is still ongoing on this event. When the investigation is completed a follow up report will be sent to the fda. Eval results: the investigation is still ongoing on this event. When the investigation is completed a follow up report will be sent to the fda. Conclusion: the investigation is still ongoing on this event. When the investigation is completed a follow up report will be sent to the fda.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1056069-2014-00001
MDR Report Key3750732
Report Source08
Date Received2014-01-31
Date of Report2014-01-07
Date Mfgr Received2014-01-20
Device Manufacturer Date2013-08-01
Date Added to Maude2014-04-16
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactKENNETH RAVENNAUGH, DIR
Manufacturer Street3545 S.W. 47TH AVE.
Manufacturer CityGAINESVILLE FL 32608
Manufacturer CountryUS
Manufacturer Postal32608
Manufacturer Phone3523360010
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSENSE BREAST 3.0T
Generic NameNA
Product CodeMOS
Date Received2014-01-31
Model Number4535-302-64711
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerINVIVO CORPORATION
Manufacturer Address3545 S.W. 47TH AVE. GAINESVILLE FL 32608 US 32608


Patients

Patient NumberTreatmentOutcomeDate
10 2014-01-31