ENDOTRACHEAL TUBE PEDIATRIC ORAL/NASAL MAGILL 3.5MM 35112 991035112

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 07 report with the FDA on 2014-04-22 for ENDOTRACHEAL TUBE PEDIATRIC ORAL/NASAL MAGILL 3.5MM 35112 991035112 manufactured by Kimberly-clark Health Care.

Event Text Entries

[4553718] Kimberly-clark received a report stating,? The nurse reported the pilot line was pulled off the endotracheal tube. The cuff deflated but the patient was not intubated. At the time of the event, the device remained in use but was eventually returned to kimberly-clark. " (b)(4).
Patient Sequence No: 1, Text Type: D, B5


[11853610] We are unable to review the device history record as no lot number was provided for the device involved in this reported event. One sample was returned. The endotracheal tube is kinked and exhibits a wavy appearance. The pilot line was detached from the sample upon receipt. The detached pilot line was viewed under magnification and a layer of adhesive was visible on the pilot line at the point in which it attaches to the endotracheal tube.. Information from this incident will be included in our product complaint and mdr trend reporting systems. Ongoing analysis of trend information is used to identify the need for additional investigations.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1033422-2014-00012
MDR Report Key3762774
Report Source07
Date Received2014-04-22
Date of Report2014-03-24
Date of Event2014-03-18
Date Mfgr Received2014-03-24
Date Added to Maude2014-06-16
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationMEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactRA/QA AE COORDINATOR
Manufacturer Street1400 HOLCOMB BRIDGE ROAD
Manufacturer CityROSWELL GA 30076
Manufacturer CountryUS
Manufacturer Postal30076
Manufacturer G1THAILAND SSCT
Manufacturer CountryTH
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameENDOTRACHEAL TUBE PEDIATRIC ORAL/NASAL MAGILL 3.5MM
Generic NameENDOTRACHEAL TUBE
Product CodeBSK
Date Received2014-04-22
Returned To Mfg2014-04-11
Model Number35112
Catalog Number991035112
Lot NumberANP
OperatorNURSE
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerKIMBERLY-CLARK HEALTH CARE
Manufacturer Address1400 HOLCOMB BRIDGE ROAD ROSWELL GA 30076 US 30076


Patients

Patient NumberTreatmentOutcomeDate
10 2014-04-22