ADULT 8.0MM ET TUBE WITH SUBGLOTTIC SUCTIONING 13222 991013222

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 07 report with the FDA on 2014-04-25 for ADULT 8.0MM ET TUBE WITH SUBGLOTTIC SUCTIONING 13222 991013222 manufactured by Kimberly-clark Health Care.

Event Text Entries

[4436144] Kimberly-clark received a report from the united kingdom stating, "when using the kc subglottic tube the pilot cuff detached from the tube, causing the cuff to deflate. This meant the patient had to be reintubated (without any apparent harm). This appeared to be spontaneous and not related to patient biting the tube. " (b)(4).
Patient Sequence No: 1, Text Type: D, B5


[11879818] We are unable to review the device history record as no lot number was provided for the device involved in this reported event. One sample device was returned. Physical appearance of sample shows pilot line was detached at the pilot line tube connection. Information from this incident will be included in our product complaint and mdr trend reporting systems. Ongoing analysis of trend information is used to identify the need for additional investigations.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number9611594-2014-00036
MDR Report Key3773987
Report Source07
Date Received2014-04-25
Date of Report2014-04-01
Date of Event2014-03-30
Date Mfgr Received2014-03-31
Date Added to Maude2014-06-18
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationMEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactRA/QA AE COORDINATOR
Manufacturer Street1400 HOLCOMB BRIDGE ROAD
Manufacturer CityROSWELL GA 30076
Manufacturer CountryUS
Manufacturer Postal30076
Manufacturer G1NOGALES 1
Manufacturer StreetCARRETERA INTN'L SALIDA NORTE #1053
Manufacturer CityMAGDALENA SONORA 84160
Manufacturer CountryMX
Manufacturer Postal Code84160
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameADULT 8.0MM ET TUBE WITH SUBGLOTTIC SUCTIONING
Generic NameENDOTRACHEAL TUBE
Product CodeBSK
Date Received2014-04-25
Returned To Mfg2014-04-10
Model Number13222
Catalog Number991013222
Lot NumberANP
OperatorOTHER HEALTH CARE PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerKIMBERLY-CLARK HEALTH CARE
Manufacturer Address1400 HOLCOMB BRIDGE ROAD ROSWELL GA 30076 US 30076


Patients

Patient NumberTreatmentOutcomeDate
10 2014-04-25

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.