IRRIGATION SET, Y-TYPE DYND19122

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2014-04-25 for IRRIGATION SET, Y-TYPE DYND19122 manufactured by Medline Industries, Inc..

Event Text Entries

[4441742] Following a surgical repair of a nicked bladder, continuous bladder irrigation was initiated. The bladder ruptured again and the surgeon indicated he was not sure if the clinician or the equipment contributed to the incident.
Patient Sequence No: 1, Text Type: D, B5


[11845607] Following the repair of a nicked bladder, the pt developed hematuria. Continuous bladder irrigation was initiated. Within 10 minutes, 3 liters of fluid had been infused into the bladder that was reported to be already weakened. Subsequently, the bladder ruptured and the pt was returned to the operating room. For surgical repair. The facility reported that the drip chamber filled completely and the nurse could not visualize the flow rate. There was no sample retained for evaluation as the incident occurred two weeks prior and was not reported at that time. We have had no other similar incidents reported to us. It is not known if the irrigation set was correctly primed with fluid prior to use. The reported amount of solution leaving the irrigation bag was significant and would be visually apparent to a clinician. The root cause appears to be clinician error. However, in an abundance of caution, this medwatch is being file.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1417592-2014-00041
MDR Report Key3788306
Report Source05,06
Date Received2014-04-25
Date of Report2014-04-23
Date of Event2014-03-10
Date Mfgr Received2014-03-24
Date Added to Maude2014-05-06
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactJULIE FINLEY
Manufacturer StreetONE MEDLINE PLACE
Manufacturer CityMUNDELEIN IL 60060
Manufacturer CountryUS
Manufacturer Postal60060
Manufacturer Phone8476434709
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameIRRIGATION SET, Y-TYPE
Product CodeEYN
Date Received2014-04-25
Catalog NumberDYND19122
Lot NumberUNK
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerMEDLINE INDUSTRIES, INC.
Manufacturer AddressMUNDELEIN IL US


Patients

Patient NumberTreatmentOutcomeDate
101. Other; 2. Required No Informationntervention 2014-04-25

© 2026 FDA.report