KLEENSPEC 590 SERIES DISP. VAGINAL SPECULA 59000

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06,08 report with the FDA on 2014-04-09 for KLEENSPEC 590 SERIES DISP. VAGINAL SPECULA 59000 manufactured by Welch Allyn.

Event Text Entries

[4380693] The customer stated that a small sized speculum had broken during a routine exam. The customer stated that the speculum was inserted into a pt and the bottom bill broke; the pt was not injured. The customer did not provide a pt identifier.
Patient Sequence No: 1, Text Type: D, B5


[11847379] Welch allyn is reporting this in an abundance of caution. The actual speculum involved in the event will not be returned to welch allyn. However, the complainant did provided a photograph of the broken speculum. The failure mode as described by the photograph has been previously investigated. The root cause of these very rare failures was determined to be related to potential impacts or loads on the shipping containers during transit or storage. No further investigation will be performed. Method: actual device not returned, photograph provided. Results: vaginal speculum. Conclusion: actual device not returned - photograph confirmed allegation.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1316463-2014-00002
MDR Report Key3793481
Report Source06,08
Date Received2014-04-09
Date of Report2014-04-03
Date of Event2014-04-03
Date Mfgr Received2014-04-03
Date Added to Maude2014-06-03
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactPEARLEY BHAMBRI, RA DIRECTOR
Manufacturer Street4341 STATE ST. RD. P.O. BOX 220
Manufacturer CitySKANEATELES FALLS NY 131530220
Manufacturer CountryUS
Manufacturer Postal131530220
Manufacturer Phone3156852568
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameKLEENSPEC 590 SERIES DISP. VAGINAL SPECULA
Product CodeHIB
Date Received2014-04-09
Model Number59000
Catalog Number59000
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerWELCH ALLYN
Manufacturer Address4341 STATE ST. RD. SKANEATELES FALLS NY 13153 US 13153


Patients

Patient NumberTreatmentOutcomeDate
10 2014-04-09

© 2026 FDA.report