CLINITEK STATUS+ ANALYZER 10379675

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,07 report with the FDA on 2014-05-06 for CLINITEK STATUS+ ANALYZER 10379675 manufactured by Siemens Healthcare Diagnotics.

Event Text Entries

[4386539] Customer reported two false negative leukocyte results on instrument, whereas reference lab results were positive. There was no report of injury due to this event.
Patient Sequence No: 1, Text Type: D, B5


[11783923] The cause for discordant leukocyte results is unknown.
Patient Sequence No: 1, Text Type: N, H10


[34966879] Visual exam by tech operations of returned instrument showed no obvious damage or undue wear. Test with status check cassette indicated normal unit functionality. Performance test of returned & comparison instrument with multistix 10sg reagent done with test solutions contrived to 5 & 10 wbc/mcl. N=12 reps each test solution, each instrument. Both returned & comparison instrument gave similar results. All strips tested at 10 wbc/mcl reported as positive, as did majority tested at 5 wbc/mcl. Instrument results agreed with visual observation of tested strips. Customer did not return reagent used in original results. Customer observation of false negative leukocyte from clinitekstatus+ s/n (b)(4) using multistix 10sg reagent cannot be confirmed.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1217157-2014-00063
MDR Report Key3793883
Report Source05,07
Date Received2014-05-06
Date of Report2014-04-10
Date of Event2014-03-26
Date Mfgr Received2014-06-25
Date Added to Maude2014-06-24
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMR. STEVEN ANDBERG
Manufacturer Street2 EDGEWATER DRIVE
Manufacturer CityNORWOOD MA 02062
Manufacturer CountryUS
Manufacturer Postal02062
Manufacturer Phone7812693655
Manufacturer G1SIEMENS HEALTHCARE DIAGNOSTICS
Manufacturer StreetNORTHERN RD
Manufacturer CitySUDBURY, SUFFOLK CO102XQ
Manufacturer CountryUK
Manufacturer Postal CodeCO10 2XQ
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCLINITEK STATUS+ ANALYZER
Generic NameCT STATUS+ ANALZER
Product CodeLJX
Date Received2014-05-06
Catalog Number10379675
OperatorOTHER HEALTH CARE PROFESSIONAL
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSIEMENS HEALTHCARE DIAGNOTICS
Manufacturer Address511 BENEDICT AVE TARRYTOWN NY 10591 US 10591


Patients

Patient NumberTreatmentOutcomeDate
10 2014-05-06

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.