MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,07 report with the FDA on 2014-05-06 for CLINITEK STATUS+ ANALYZER 10379675 manufactured by Siemens Healthcare Diagnotics.
[4386539]
Customer reported two false negative leukocyte results on instrument, whereas reference lab results were positive. There was no report of injury due to this event.
Patient Sequence No: 1, Text Type: D, B5
[11783923]
The cause for discordant leukocyte results is unknown.
Patient Sequence No: 1, Text Type: N, H10
[34966879]
Visual exam by tech operations of returned instrument showed no obvious damage or undue wear. Test with status check cassette indicated normal unit functionality. Performance test of returned & comparison instrument with multistix 10sg reagent done with test solutions contrived to 5 & 10 wbc/mcl. N=12 reps each test solution, each instrument. Both returned & comparison instrument gave similar results. All strips tested at 10 wbc/mcl reported as positive, as did majority tested at 5 wbc/mcl. Instrument results agreed with visual observation of tested strips. Customer did not return reagent used in original results. Customer observation of false negative leukocyte from clinitekstatus+ s/n (b)(4) using multistix 10sg reagent cannot be confirmed.
Patient Sequence No: 1, Text Type: N, H10
Report Number | 1217157-2014-00063 |
MDR Report Key | 3793883 |
Report Source | 05,07 |
Date Received | 2014-05-06 |
Date of Report | 2014-04-10 |
Date of Event | 2014-03-26 |
Date Mfgr Received | 2014-06-25 |
Date Added to Maude | 2014-06-24 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Reporter Occupation | OTHER HEALTH CARE PROFESSIONAL |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 0 |
Event Location | 0 |
Manufacturer Contact | MR. STEVEN ANDBERG |
Manufacturer Street | 2 EDGEWATER DRIVE |
Manufacturer City | NORWOOD MA 02062 |
Manufacturer Country | US |
Manufacturer Postal | 02062 |
Manufacturer Phone | 7812693655 |
Manufacturer G1 | SIEMENS HEALTHCARE DIAGNOSTICS |
Manufacturer Street | NORTHERN RD |
Manufacturer City | SUDBURY, SUFFOLK CO102XQ |
Manufacturer Country | UK |
Manufacturer Postal Code | CO10 2XQ |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | CLINITEK STATUS+ ANALYZER |
Generic Name | CT STATUS+ ANALZER |
Product Code | LJX |
Date Received | 2014-05-06 |
Catalog Number | 10379675 |
Operator | OTHER HEALTH CARE PROFESSIONAL |
Device Age | DA |
Device Eval'ed by Mfgr | Y |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | SIEMENS HEALTHCARE DIAGNOTICS |
Manufacturer Address | 511 BENEDICT AVE TARRYTOWN NY 10591 US 10591 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2014-05-06 |