5.0MM X 10FT FIBEROPTIC LIGHT CABLE 0233065010

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 07 report with the FDA on 2014-05-06 for 5.0MM X 10FT FIBEROPTIC LIGHT CABLE 0233065010 manufactured by Stryker Endoscopy-san Jose.

Event Text Entries

[4476199] It was reported that the light cable was turned causing it to touch the patient's abdomen, which allegedly caused a surface burn.
Patient Sequence No: 1, Text Type: D, B5


[11882073] Additional information will be provided once the investigation has been completed.
Patient Sequence No: 1, Text Type: N, H10


[19972787] It was reported that the light cable was turned causing it to touch the patient's abdomen, which allegedly caused a surface burn.
Patient Sequence No: 1, Text Type: D, B5


[20305956] The device was not returned to stryker endo for investigation. A detailed investigation cannot be performed without the unit, this complaint will be closed as unit not received and based on a probable root cause. In the event that if the unit is returned to stryker endo, we will reopen the investigation for further analysis. Based on previous investigations and the risk document, a probable root cause may be too much heat generated from the light source. The product was not returned for investigation and the reported failure mode could not be confirmed. The failure mode will be monitored for future reoccurrence.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number0002936485-2014-00308
MDR Report Key3793982
Report Source07
Date Received2014-05-06
Date of Report2014-03-18
Date of Event2014-03-14
Date Mfgr Received2014-03-18
Date Added to Maude2014-05-06
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMR. THOMAS SHAFER
Manufacturer Street5900 OPTICAL COURT
Manufacturer CitySAN JOSE CA 95138
Manufacturer CountryUS
Manufacturer Postal95138
Manufacturer Phone4087542000
Manufacturer G1STRYKER ENDOSCOPY-SAN JOSE
Manufacturer Street5900 OPTICAL COURT
Manufacturer CitySAN JOSE CA 95138
Manufacturer CountryUS
Manufacturer Postal Code95138
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand Name5.0MM X 10FT FIBEROPTIC LIGHT CABLE
Generic NameLIGHT, SURGICAL, FLOOR STANDING
Product CodeFSS
Date Received2014-05-06
Catalog Number0233065010
Lot NumberUNK
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerSTRYKER ENDOSCOPY-SAN JOSE
Manufacturer Address5900 OPTICAL COURT SAN JOSE CA 95138 US 95138


Patients

Patient NumberTreatmentOutcomeDate
10 2014-05-06