PROGENY VANTAGE 60-A0001 V5000

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 08 report with the FDA on 2014-05-08 for PROGENY VANTAGE 60-A0001 V5000 manufactured by Midmark Corporation.

Event Text Entries

[4364971] A dental assistant of a dental office reported to a dealer field technician who then reported to midmark that the column of a vantage x-ray system, serial number (b)(4), was observed to move up without the movement control button being pressed. The dealer technician inspected the machine and reported that he double tapped on the left membrane switch to make the unit move down, and the unit kept moving downward without the membrane switch being pressed. The machine has a movement control membrane switch on both sides of the patient table. The malfunction only happened when the membrane switch on the left hand side was used. Midmark confirmed with the dental office that there was no injury.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1423380-2014-00003
MDR Report Key3800655
Report Source08
Date Received2014-05-08
Date of Report2014-05-06
Date of Event2014-04-08
Date Mfgr Received2014-04-09
Device Manufacturer Date2011-05-24
Date Added to Maude2014-06-25
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationDENTAL ASSISTANT
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactPENG LIN
Manufacturer Street675 HEATHROW DRIVE
Manufacturer CityLINCOLNSHIRE IL 60065
Manufacturer CountryUS
Manufacturer Postal60065
Manufacturer Phone8474159787
Manufacturer G1MIDMARK CORPORATION
Manufacturer Street675 HEATHROW DRIVE
Manufacturer CityLINCOLNSHIRE IL 60069
Manufacturer CountryUS
Manufacturer Postal Code60069
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NamePROGENY VANTAGE
Generic NameSYSTEM,X-RAY,EXTRAORAL SOURCE,DIGITAL
Product CodeMUH
Date Received2014-05-08
Model Number60-A0001
Catalog NumberV5000
OperatorDENTAL ASSISTANT
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerMIDMARK CORPORATION
Manufacturer Address675 HEATHROW DRIVE LINCOLNSHIRE IL 60069 US 60069


Patients

Patient NumberTreatmentOutcomeDate
10 2014-05-08

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.