VERSACELL SAMPLE MANAGEMENT SYSTEM 030102

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2014-05-08 for VERSACELL SAMPLE MANAGEMENT SYSTEM 030102 manufactured by Siemens Healthcare Diagnostics Inc..

Event Text Entries

[19021827] Three patient sample tubes were dropped by a versacell sample management system while being transferred to an immulite instrument for testing. The samples spilled and were cleaned by the laboratory staff, who wore personal protective equipment. The samples could not be processed by the instrument and new samples had to be obtained from the patients. The spilled contents from the dropped sample tubes did not contaminate any other samples scheduled for testing. There are no known reports of patient intervention or adverse health consequences due to the sample tubes being dropped by the versacell sample management system prior to being sampled by the instrument.
Patient Sequence No: 1, Text Type: D, B5


[19049354] The customer contacted the siemens customer care center (ccc). During troubleshooting with the customer, it was determined that the versacell sample management system grippers appeared to be working as expected. There were no obstructions and all o-rings were in place. A siemens customer service engineer (cse) was dispatched to the customer site. The cse replaced the x-direction server motor printed circuit board and adjusted the arm alignment. The cause of the dropped sample tubes is unknown. The versacell sample management system is performing according to specifications. No further evaluation of this device is required.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2247117-2014-00025
MDR Report Key3801407
Report Source05,06
Date Received2014-05-08
Date of Report2014-04-15
Date of Event2014-04-15
Date Mfgr Received2014-04-15
Date Added to Maude2014-06-25
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactCASSANDRA KOCSIS
Manufacturer Street511 BENEDICT AVE
Manufacturer CityTARRYTOWN NY 10591
Manufacturer CountryUS
Manufacturer Postal10591
Manufacturer Phone9145242687
Manufacturer G1SIEMENS HEALTHCARE DIAGNOSTICS INC.
Manufacturer Street62 FLANDERS BARTLEY RD
Manufacturer CityFLANDERS NJ 07836
Manufacturer CountryUS
Manufacturer Postal Code07836
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameVERSACELL SAMPLE MANAGEMENT SYSTEM
Generic NameSAMPLE MANAGEMENT SYSTEM
Product CodeDGC
Date Received2014-05-08
Model NumberVERSACELL SAMPLE MANAGEMENT SYSTEM
Catalog Number030102
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerSIEMENS HEALTHCARE DIAGNOSTICS INC.
Manufacturer Address62 FLANDERS BARTLEY RD FLANDERS NJ 07836 US 07836


Patients

Patient NumberTreatmentOutcomeDate
10 2014-05-08

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.