YASARGIL TI PERM MINI-CLIP CVD 3.9MM FT714T

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,07 report with the FDA on 2014-04-16 for YASARGIL TI PERM MINI-CLIP CVD 3.9MM FT714T manufactured by Aesculap Ag & Co. Kg.

Event Text Entries

[4383743] Country of complaint: (b)(6). The surgeon clipped the operative site by ft714t and ft742t in tandem. The surgeon found a clip had slipped off when the angiography image was reviewed. Which clip (ft714t or ft742t) slipped off is under investigation. State of brain aneurysm: dome diameter - 4mm. Neck diameter - 2mm. Device 1 of 2.
Patient Sequence No: 1, Text Type: D, B5


[11891469] Manufacturing site evaluation: clips received show traces of wear which may be a sign of multiple use, or improper use of application forceps. There are no indications of manufacturing or material defects. The cause of the failure can not be determined.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2916714-2014-00244
MDR Report Key3810475
Report Source01,07
Date Received2014-04-16
Date of Report2014-04-16
Date of Event2014-03-07
Date Mfgr Received2014-03-17
Device Manufacturer Date2013-10-01
Date Added to Maude2014-05-28
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMICHELLE LINK
Manufacturer Street615 LAMBERT POINTE DRIVE
Manufacturer CityHAZELWOOD MO 63042
Manufacturer CountryUS
Manufacturer Postal63042
Manufacturer Phone3145515938
Manufacturer G1AESCULAP AG&CO KG
Manufacturer StreetPO BOX 40
Manufacturer CityTUTTLINGEN 78501
Manufacturer CountryGM
Manufacturer Postal Code78501
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameYASARGIL TI PERM MINI-CLIP CVD 3.9MM
Generic NameCEREBRO VASCULAR CLIPS
Product CodeHCH
Date Received2014-04-16
Model NumberFT714T
Catalog NumberFT714T
Lot Number51972764
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerAESCULAP AG & CO. KG
Manufacturer AddressTUTTLINGEN DE 78532 US 78532


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2014-04-16

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.