COBAS AMPLICOR ANALYZER 21045156001

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2014-05-15 for COBAS AMPLICOR ANALYZER 21045156001 manufactured by Roche Diagnostics Ltd..

Event Text Entries

[4525852] A customer in the united states filed a complaint case as it was identified that the resuspension tip tubing on the cobas amplicor analyzer was accidentally inverted during periodic maintenance of their system, which was performed by the local field service engineer. This inversion was identified as the customer's cobas amplicor analyzer was leaking when they were performing cobas ampliscreen hcv test, v2. 0 runs on the system. Upon identification of the issue, the customer's system was repaired and the resuspension tip tubing was re-installed appropriately by the local field service engineer. All donor samples that generated non-reactive hcv results with the cobas ampliscreen hcv test, v2. 0 during the timeframe of when the resuspension tip tubing was inverted ((b)(6) 2014) were subjected to repeat testing; a total of 30 donor samples were repeated tested. One (1) donor out of the 30 donor samples tested generated discrepant results between initial testing (non-reactive hcv result) and repeat testing (reactive hcv result); the same lots of cobas ampliscreen hcv test, v2. 0 were used for both test runs. The serology result for this donor could not be confirmed. It is unknown if the initial non-reactive hcv result was reported and if the donor sample was released; this information could not be confirmed as the customer has not been willing to provide additional details surrounding this situation. The true diagnosis of the donor is unknown. The customer did not repeat test those donor samples that generated reactive hcv results as these results were aligned with positive serology results. This mdr is specific to the cobas amplicor analyzer as the inversion of the resuspension tip tubing could potentially result in inadequate dispension of wash reagent to the reaction cups used during the viral detection.
Patient Sequence No: 1, Text Type: D, B5


[11890376] Result: failure to service/maintain according to the manufacturer's recommendation. Conclusion: device failure related to maintenance 24 off label, unapproved, or contradicted use. The investigation of this issue indicated that the local field service engineer had inverted the resuspension tip tubing during periodic maintenance of the customer's cobas amplicor analyzer; the service documentation does provide adequate instruction on how to replace resuspension tip tubing and specifies that the engineer must verify that the "fittings correspond to the correct clips at the wash arm" as well as provides an image to assist with this verification. The inversion of the resuspension tip tubing could potentially result in inadequate dispension of wash reagent to the reaction cups used during viral detection; inappropriate washing may impact test results, and this has not been evaluated by roche nor confirmed for the current case or for the donor sample that generated discrepant hcv results as the donor sample was not returned for further evaluation. There is no indication of a product issue with the cobas ampliscreen hcv test, v2. 0 kits indicated to be used at the site as they have been determined to be within specification and performing as intended. It is unknown if the initial non-reactive hcv result was reported and if the donor sample was released; this information could not be confirmed as the customer has not been willing to provide additional details surrounding this situation. The true diagnosis of the donor is unknown. (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[34907905] Within the initial mdr 2243471-2014-00015 filed, it was indicated that it was unknown whether the sample that generated the discrepant results was released for transfusion. It was later clarified that the sample was appropriately discarded and not transfused. (b)(4).
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2243471-2014-00015
MDR Report Key3813623
Report Source05
Date Received2014-05-15
Date of Report2014-05-19
Date of Event2014-04-07
Date Mfgr Received2014-05-19
Date Added to Maude2014-06-26
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactVINCENT STAGNITTO
Manufacturer Street1080 US HWY 202S
Manufacturer CityBRANCHBURG NJ 088763733
Manufacturer CountryUS
Manufacturer Postal088763733
Manufacturer Phone9082537569
Manufacturer G1ROCHE DIAGNOSTICS LTD.
Manufacturer StreetFORRENSTRASSE
Manufacturer CityROTKREUZ, CH-6343
Manufacturer CountrySZ
Manufacturer Postal CodeCH-6343
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCOBAS AMPLICOR ANALYZER
Generic NameANALYZER, CHEMISTY, MICRO, FOR CLINICAL USE
Product CodeJJF
Date Received2014-05-15
Catalog Number21045156001
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS LTD.
Manufacturer AddressFORRENSTRASSE ROTKREUZ, CH-6343 SZ CH-6343


Patients

Patient NumberTreatmentOutcomeDate
10 2014-05-15

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.