3500 XL DX GENETIC ANALYZER CS2 4461444

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 04,07 report with the FDA on 2014-03-27 for 3500 XL DX GENETIC ANALYZER CS2 4461444 manufactured by Life Technologies Holdings Pte. Ltd..

Event Text Entries

[19022828] A customer reported that the 3500xl dx genetic analyzer cs2 (cat no 4461444, serial no (b)(4)) stopped in the middle of an assay run due to the presence air bubbles. No patient involvement reported. The company event tracking number is (b)(4).
Patient Sequence No: 1, Text Type: D, B5


[19132987] Device is an instrument and is not implanted/explanted. The instrument was investigated on site. The investigation determined the cause of the air bubbles was due to a leaking pump. The pump was replaced and spatial calibration was run after the repair. The instrument passed required spatial calibration specifications. Device intended use: the applied biosystems 3500xl dx genetic analyzer cs2 with 3500 dx series software 2011 (vl) are in vitro diagnostic devices intended for the sequencing (detection and identification) of fluorescently labeled deoxyribonucleic acid (dna) by capillary electrophoresis. The applied biosystems 3500xl dx sequencing assays specifying their use and only by technologists trained in laboratory techniques, procedures, and use of the analyzer.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3003673482-2014-00007
MDR Report Key3815366
Report Source04,07
Date Received2014-03-27
Date of Report2014-01-20
Date of Event2014-01-20
Date Mfgr Received2014-01-20
Device Manufacturer Date2013-10-01
Date Added to Maude2014-06-16
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactKELLI TANZELLA
Manufacturer Street3175 STALEY ROAD, GRAND ISLAND
Manufacturer CityNEW YORK NY 14072
Manufacturer CountryUS
Manufacturer Postal14072
Manufacturer Phone7167743122
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand Name3500 XL DX GENETIC ANALYZER CS2
Generic NameDNA GENETIC ANALYZER
Product CodePCA
Date Received2014-03-27
Catalog Number4461444
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerLIFE TECHNOLOGIES HOLDINGS PTE. LTD.
Manufacturer AddressSINGAPORE SN


Patients

Patient NumberTreatmentOutcomeDate
10 2014-03-27

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.