TRIAGE + TCA 25 TEST 92000

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2014-02-18 for TRIAGE + TCA 25 TEST 92000 manufactured by Alere San Diego Inc.

Event Text Entries

[19973232] Customer reported potential (b)(6) coc on two samples with triage + tca 25 test vs. Confirmatory testing. Two samples were taken from a female patient who delivered a baby by c-section. Samples were collected (b)(6) 2014. Ran the first sample before c-section, collected at 1630. Re-collected a second sample and ran after c-section, collected at 1810. The patient tested positive for coc with the triage + tca 25 test. Coc did not test positive on the confirmation testing that was performed.
Patient Sequence No: 1, Text Type: D, B5


[20261876] No discrepant or positive results were observed with any devices tested. No device issues or errors were observed with any device tested. No patient sample was returned for testing. Patient sample interference cannot be ruled out as a potential cause for discrepant results. No product deficiency was established. No corrective action required at this time.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2027969-2014-00134
MDR Report Key3822927
Report Source05
Date Received2014-02-18
Date of Report2014-02-12
Date of Event2014-01-30
Date Mfgr Received2014-02-12
Date Added to Maude2014-06-09
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactYA-LING KING
Manufacturer Street9975 SUMMERS RIDGE RD
Manufacturer CitySAN DIEGO CA 92121
Manufacturer CountryUS
Manufacturer Postal92121
Manufacturer Phone8588052084
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTRIAGE + TCA 25 TEST
Generic NameDRUG OF ABUSE TEST
Product CodeDKZ
Date Received2014-02-18
Model Number92000
Lot Number326218
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerALERE SAN DIEGO INC
Manufacturer AddressSAN DIEGO CA US


Patients

Patient NumberTreatmentOutcomeDate
10 2014-02-18

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.