3500 XL DX GENETIC ANALYZER CS2 4461444

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 04,07 report with the FDA on 2014-03-06 for 3500 XL DX GENETIC ANALYZER CS2 4461444 manufactured by Life Technologies Holdings Pte. Ltd.

Event Text Entries

[4499252] A customer reported there was spatial calibration failure in the instrument, 3500 xl dx genetic analyzer cs2 (cat. No. 4461444, serial no. (b)(4)). No patient involvement reported.
Patient Sequence No: 1, Text Type: D, B5


[11745450] Device is an instrument and is not implanted/explanted. The instrument was investigated on site. It was discovered that the array door was binding, preventing the laser to pass through capillary. The door/spring was disassembled and lubricated. It was then reassembled and installed. Several spatial calibrations were run after the repair and the instrument passed required specifications. Device intended use: the applied biosystems 3500xl dx genetic analyzer cs2 with 3500 dx series software 2011 (vl) are invitro diagnostic devices intended for the sequencing (detection and identification) of fluorescently-labeled deoxyribonucleic acid (dna) by capillary electrophoresis. The applied biosystems 3500xl dx genetic analyzer cs2 with 3500 dx sequencing assays specifying their use and only by technologists trained in laboratory techniques, procedures and use of the analyzer.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3003673482-2014-00004
MDR Report Key3831934
Report Source04,07
Date Received2014-03-06
Date of Report2014-02-07
Date of Event2014-02-07
Date Mfgr Received2014-02-07
Device Manufacturer Date2013-03-01
Date Added to Maude2014-06-19
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactKELLI TANZELLA
Manufacturer Street3175 STALEY ROAD
Manufacturer CityGRAND ISLAND NY 14072
Manufacturer CountryUS
Manufacturer Postal14072
Manufacturer Phone7167743122
Manufacturer G1LIFE TECHNOLOGIES HOLDINGS PTE.LTD
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand Name3500 XL DX GENETIC ANALYZER CS2
Generic NamePCA
Product CodePCA
Date Received2014-03-06
Catalog Number4461444
Lot Number25800-221
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerLIFE TECHNOLOGIES HOLDINGS PTE. LTD
Manufacturer AddressSINGAPORE SN


Patients

Patient NumberTreatmentOutcomeDate
10 2014-03-06

© 2026 FDA.report