PLASTIBELL HOSP BULK 1.2 CM D 9212

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2014-05-05 for PLASTIBELL HOSP BULK 1.2 CM D 9212 manufactured by Hollister Inc..

Event Text Entries

[4393194] The unit broke in half during a circumcision procedure. The physician was tying the string and the plastibell popped off and hit the physician in the chest. When the physician picked the plastibell off the floor, he noticed that the string had tightened around the patient's penis, cutting off the tip. The physician cut off the string from the patient's penis. The patient had to have an add'l surgery. The hospital has been using the product since november 2007. The physician has been using the product for 10 years. A review of the device history for this particular model and lot number was performed, and no abnormalities were discovered. A review of complaint history for this product reveals no similar incidents.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1422443-2014-00001
MDR Report Key3832122
Report Source05
Date Received2014-05-05
Date of Report2014-05-02
Date of Event2014-04-03
Date Mfgr Received2014-04-04
Device Manufacturer Date2013-04-06
Date Added to Maude2014-05-28
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationRISK MANAGER
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. JERI HILL
Manufacturer Street1931 NORMAN DR.
Manufacturer CityWAUKEGAN IL 60085
Manufacturer CountryUS
Manufacturer Postal60085
Manufacturer Phone8475735947
Manufacturer G1HOLLISTER INC.
Manufacturer Street1502 EAST LAHARPE
Manufacturer CityKIRKSVILLE MO 63501
Manufacturer CountryUS
Manufacturer Postal Code63501
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NamePLASTIBELL HOSP BULK 1.2 CM
Generic NamePLASTIBELL CIRCUMCISION DEVICE
Product CodeFHG
Date Received2014-05-05
Model NumberD 9212
Catalog NumberD 9212
Lot Number3D06
Device Expiration Date2018-04-30
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerHOLLISTER INC.
Manufacturer Address1502 EAST LAHARPE KIRKSVILLE MO 63501 US 63501


Patients

Patient NumberTreatmentOutcomeDate
101. Hospitalization 2014-05-05

© 2026 FDA.report