STAINLESS STEEL RULER, 30 0MM SR

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2014-02-10 for STAINLESS STEEL RULER, 30 0MM SR manufactured by Integra Burlington Ma, Inc.

Event Text Entries

[17574717] It was reported that the numbers at 10 cm intervals of the ruler were filled with red paint while the rest of the numbers were black. The red paint flakes off and contaminates other parts like the cannulas. The surgeon spent some time in a recent deep brain stimulation (dbs) case cleaning paint flakes from inside a cannula. Additional information was requested and on (b)(4) 2014, the following was received: the product problem occurred on (b)(6) 2013 during the preparation of dbs electrode for insertion to target depth after completion of neuron signal mapping (mer). The customer explained that using the axon instruments system, a guide tube is placed on the ruler, after having set the slider on the ruler to the target depth. Then, an adhesive paper tab is wrapped around the upper end of the guide tube to act as a temporary depth stop 5 mm above target. The dbs electrode is inserted into the guide tube and fixed in place by the adhesive tab extending past the upper end of the guide tube. In this case, the surgeon saw red paint flakes on the guide tube sometime during the above process and stopped to clean them off the surface and flush the lumen with saline. When the product problem occurred, the (b)(6) male patient was awake but sedated. Surgical delay was under 10 minutes. There was no known adverse consequence as a result of the surgical delay.
Patient Sequence No: 1, Text Type: D, B5


[17714430] The device involved in the reported incident is not expected to be received for evaluation. An investigation has been initiated based upon the reported information.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1222895-2014-00002
MDR Report Key3832315
Report Source05,06
Date Received2014-02-10
Date of Report2014-01-20
Date of Event2013-12-09
Date Mfgr Received2014-01-20
Date Added to Maude2014-06-16
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactROWENA BUNUAN
Manufacturer Street315 ENTERPRISE DRIVE
Manufacturer CityPLAINSBORO NJ 08536
Manufacturer CountryUS
Manufacturer Postal08536
Manufacturer Phone6099362393
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSTAINLESS STEEL RULER, 30 0MM
Generic NameNA
Product CodeFTY
Date Received2014-02-10
Catalog NumberSR
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerINTEGRA BURLINGTON MA, INC
Manufacturer AddressBURLINGTON MA 01803 US 01803


Patients

Patient NumberTreatmentOutcomeDate
10 2014-02-10