YASARGIL TI PERM STD-CLIP STR 11MM FT760T

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,07 report with the FDA on 2014-03-12 for YASARGIL TI PERM STD-CLIP STR 11MM FT760T manufactured by Aesculap Ag & Co. Kg.

Event Text Entries

[4407546] Country of complaint: (b)(6). Clips slipped out 3-4 days post surgery. Revision surgery is necessary due to the slipped clip.
Patient Sequence No: 1, Text Type: D, B5


[11806466] U. S. Reporting agent notified on (b)(4) 2014. Manufacturing site eval: jaw is no longer opaque and no longer closes completely. Complete surface is severely scratched. Various pressure marks in the shank and spring range. Significant closing force drop of 173 g to 145g. Conclusion: the scratched surface and the pressure marks indicate a multiple use. The pressure marks in the shank suggest that the clip was slanted in the applying forceps. Improper treatment has caused the overloading of the clip and thus the spring and the clamping force is adversely affected. Such is delivery state can be ruled out at 100%.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2916714-2014-00130
MDR Report Key3845507
Report Source01,07
Date Received2014-03-12
Date of Report2014-03-07
Date of Event2014-01-24
Date Mfgr Received2014-01-29
Device Manufacturer Date2013-02-01
Date Added to Maude2014-06-20
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMICHELLE LINK
Manufacturer Street615 LAMBERT POINTE DR
Manufacturer CityHAZELWOOD MO 63042
Manufacturer CountryUS
Manufacturer Postal63042
Manufacturer Phone3145515938
Manufacturer G1AESCULAP AG & CO. KG
Manufacturer StreetPO BOX 40
Manufacturer CityTUTTLINGEN 78501
Manufacturer CountryGM
Manufacturer Postal Code78501
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameYASARGIL TI PERM STD-CLIP STR 11MM
Product CodeHCH
Date Received2014-03-12
Model NumberFT760T
Catalog NumberFT760T
Lot Number51909695
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerAESCULAP AG & CO. KG
Manufacturer AddressTUTTLINGEN DE 78532 DE 78532


Patients

Patient NumberTreatmentOutcomeDate
10 2014-03-12

© 2026 FDA.report