AU5400 CLINICAL CHEMISTRY ANALYZER AU5400-30 N1464600

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2014-06-04 for AU5400 CLINICAL CHEMISTRY ANALYZER AU5400-30 N1464600 manufactured by Beckman Coulter.

Event Text Entries

[4686060] The customer reported obtaining a potassium result with an exclamation point flag, "! ", for one patient on an au5400 clinical chemistry analyzer. The flag indicates the analyzer was unable to calculate concentration. The patient sample was rerun and a result was obtained without flags. There were no patient results released from the laboratory and there was no change to patient treatment in connection with this event.
Patient Sequence No: 1, Text Type: D, B5


[11958072] A beckman coulter (bec) field service engineer (fse) was dispatched to the customer's site. The fse found a rusty grounding point at the bottom of the ion selective electrode (ise) unit. The fse stated that because ise unit was not grounded, it would pick up extraneous noise. The fse cleaned the grounding point and performed a precision study. There were no further reports of issues.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number9612296-2014-00084
MDR Report Key3846742
Report Source05,06
Date Received2014-06-04
Date of Report2014-05-13
Date of Event2014-05-12
Date Mfgr Received2014-05-13
Device Manufacturer Date2003-07-10
Date Added to Maude2014-07-03
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMS. DUNG NGUYEN
Manufacturer Street250 S. KRAEMER BLVD.
Manufacturer CityBREA CA 92821
Manufacturer CountryUS
Manufacturer Postal92821
Manufacturer Phone7149614941
Manufacturer G1BECKMAN COULTER MISHIMA K.K.
Manufacturer Street454-32 HIGASHINO,NAGAIZUMI-CHO
Manufacturer CitySUNTO-GUN, SCHIZUOKA, JP-NOTA 411-0931
Manufacturer CountryJA
Manufacturer Postal Code411-0931
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameAU5400 CLINICAL CHEMISTRY ANALYZER
Generic NameANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE
Product CodeMZV
Date Received2014-06-04
Model NumberAU5400-30
Catalog NumberN1464600
Lot NumberNA
ID NumberSW VERSION 5.5
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBECKMAN COULTER
Manufacturer Address250 S. KRAEMER BLVD. BREA CA 92821 US 92821


Patients

Patient NumberTreatmentOutcomeDate
10 2014-06-04

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.