HAWKINS II BLN (BREAST LOCALIZATION NEEDLE) 242050

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2014-05-28 for HAWKINS II BLN (BREAST LOCALIZATION NEEDLE) 242050 manufactured by Argon Medical Devices Inc..

Event Text Entries

[4616533] Dr (b)(6), proceeded to deploy the wire through the end hole and noticed that he did not feel it lock into place. He then realized that the end of the hook broke off from the wire. He then placed another needle/wire into the breast. The patient needed to have more tissue removed to make sure the broken piece of wire from the hawkins ii needle was excised with the abnormality.
Patient Sequence No: 1, Text Type: D, B5


[12163479] A review of the device history record did not reveal any related deviations or abnormalities for the complaint lot number 11061821. There was one other similar complaint for this product from the same doctor. Before these two complaints, there were no other prior similar complaints for lot 11061821 or 11063213 which both contained the same sub (wire) lot. The root cause of the breakage of the breast localization needle was not determined. The returned needle was broken. Complaints will continue to be monitored for trends.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1625425-2014-00017
MDR Report Key3854900
Report Source06
Date Received2014-05-28
Date of Report2014-05-27
Date of Event2014-04-24
Date Mfgr Received2014-04-28
Device Manufacturer Date2014-03-06
Date Added to Maude2014-06-10
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactGAIL SMITH
Manufacturer Street1445 FLAT CREEK RD.
Manufacturer CityATHENS TX 75751
Manufacturer CountryUS
Manufacturer Postal75751
Manufacturer Phone9722102562
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameHAWKINS II BLN (BREAST LOCALIZATION NEEDLE)
Product CodeGDM
Date Received2014-05-28
Returned To Mfg2014-05-02
Catalog Number242050
Lot Number11061821
Device Expiration Date2019-03-01
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerARGON MEDICAL DEVICES INC.
Manufacturer AddressATHENS TX US


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2014-05-28

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.