BIOTRUE MULTI-PURPOSE SOLUTION

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 04 report with the FDA on 2014-04-21 for BIOTRUE MULTI-PURPOSE SOLUTION manufactured by Bausch & Lomb.

Event Text Entries

[20972282] As a result of a consumer complaint, retain samples of tank lot jk3573 were tested and confirmed to have been out of specification (undetectable) for one of the two disinfectants: polyquaternium-1 (pq-1). The pq-a specification is (b)(4). Pq-1 id testing did show pq-1 to be present. Due to the remaining 6 months of the 24 month shelf life for finished product filled from tank jk3573, the decision was made to recall the associated finished product. There is no health hazard concern as the other disinfectant was shown to be at the proper level and shown through additional laboratory testing to provide disinfectant coverage. There is an on-going investigation into the root cause for the out of specification result.
Patient Sequence No: 1, Text Type: D, B5


[21153207] The returned sample was tested and found to eb out of specification for one of the two disinfectants; polyquaternium-1 (pq-1) was not detected.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1313525-2014-00005
MDR Report Key3855772
Report Source04
Date Received2014-04-21
Date of Report2014-04-14
Date Mfgr Received2014-04-14
Device Manufacturer Date2012-10-01
Date Added to Maude2014-06-23
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer Street1400 NORTH GOODMAN ST.
Manufacturer CityROCHESTER NY 14609
Manufacturer CountryUS
Manufacturer Postal14609
Manufacturer Phone5853388549
Manufacturer G1BAUSCH & LOMB, INC.
Manufacturer Street8507 PELHAM ROAD
Manufacturer CityGREENVILLE SC 29615
Manufacturer CountryUS
Manufacturer Postal Code29615
Single Use3
Remedial ActionRC
Previous Use Code3
Removal Correction Number1032500-04/21/2014-001-R
Event Type3
Type of Report3

Device Details

Brand NameBIOTRUE MULTI-PURPOSE SOLUTION
Generic NameCONTACT LENS MULTI-PURPOSE SOLUTION
Product CodeLYL
Date Received2014-04-21
Returned To Mfg2014-02-03
Lot NumberGK2127
Device Expiration Date2014-10-31
OperatorLAY USER/PATIENT
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBAUSCH & LOMB
Manufacturer AddressROCHESTER NY US


Patients

Patient NumberTreatmentOutcomeDate
10 2014-04-21