REUSABLE ELECTRODE EXTENDER SAME 138026

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 1993-02-02 for REUSABLE ELECTRODE EXTENDER SAME 138026 manufactured by Birtcher Medical Systems, Inc..

Event Text Entries

[2244] Forty-eight year old woman in good condition to or for left breast revision reconstruction. During the procedure while the physician was caauterizating using the extension, a spark ocurred causing a burn at the cautery szite. The burned area ws excised. The extender was examined for cracks in the insulation and none was founddevice not labeled for single use. Patient medical status prior to event: satisfactory condition. There was not multiple patient involvement. Invalid data - on device service/maintenance. No data - regarding date last serviced. Service provided by: invalid data. Invalid data - service records availability. No imminent hazard to public health claimed. Device used as labeled/intended. Device was evaluated after the event. Method of evaluation: actual device involved in incident was evaluated, other. Results of evaluation: none or unknown. Conclusion: none or unknown. Certainty of device as cause of or contributor to event: maybe. Corrective actions: device permanently removed from service, other. Invalid data - on device destroyed/disposed of status.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3856
MDR Report Key3856
Date Received1993-02-02
Date of Report1992-12-01
Date of Event1992-11-27
Date Facility Aware1992-11-27
Report Date1992-12-01
Date Reported to FDA1992-12-01
Date Reported to Mfgr1992-12-01
Date Added to Maude1993-05-06
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional0
Initial Report to FDA0
Report to FDA3
Event Location3
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Details

Brand NameREUSABLE ELECTRODE EXTENDER
Generic NameELECTROCAUTERY EXTENSION
Product CodeHIM
Date Received1993-02-02
Model NumberSAME
Catalog Number138026
Lot NumberN/A
ID NumberN/A
OperatorOTHER HEALTH CARE PROFESSIONAL
Device AvailabilityY
Implant FlagN
Device Sequence No1
Device Event Key3596
ManufacturerBIRTCHER MEDICAL SYSTEMS, INC.


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 1993-02-02

© 2026 FDA.report