EASYSTAND STRAPSTAND 2000

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 04 report with the FDA on 2014-04-25 for EASYSTAND STRAPSTAND 2000 manufactured by Altimate Medical, Inc..

Event Text Entries

[4617498] Received a call from a woman who purchased a strapstand in (b)(6) 2012 for her husband. She said her and a helper were putting him in the unit when one side of the strap tore away from the padded seat body of the xl adjustable seat strap on an easystand strapstand. The user was not injured but startled when it occurred. Both of them were on each side of him and he fell back into his wheelchair, which is positioned up to the standing frame, so it's available when the user goes back to the seated position.
Patient Sequence No: 1, Text Type: D, B5


[12161344] The adjustable seat strap was returned to altimate and reviewed. Upon review, it was noted that the inserted strap in the seat missed the last row of stitching. This appears to have been caused by inconsistent manufacturing processes at the supplier who manufactures these straps. Altimate medical has updated the product specifications to add add'l detailed info to the product specifications for this component and has communicated with the supplier regarding their quality control process. Altimate has also implemented a receiving inspection process on this component in addition to the current inspection process performed during the assembly and shipping process.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2183634-2014-00002
MDR Report Key3856174
Report Source04
Date Received2014-04-25
Date of Report2014-03-26
Date of Event2014-03-26
Date Mfgr Received2014-03-26
Device Manufacturer Date2012-08-01
Date Added to Maude2014-06-20
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactSTACEY FRANK
Manufacturer Street262 WEST FIRST ST.
Manufacturer CityMORTON MN 56270
Manufacturer CountryUS
Manufacturer Postal56270
Manufacturer Phone5076976393
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameEASYSTAND STRAPSTAND
Generic Name890.5370
Product CodeION
Date Received2014-04-25
Returned To Mfg2014-04-07
Model NumberSTRAPSTAND
Catalog Number2000
Lot NumberNA
OperatorLAY USER/PATIENT
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerALTIMATE MEDICAL, INC.
Manufacturer Address262 WEST FIRST ST. MORTON MN 56270 US 56270


Patients

Patient NumberTreatmentOutcomeDate
10 2014-04-25

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.