TRANSCUTANEOUS BLOOD GAS MONITOR TCM4 SERIES 391-880

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2014-05-29 for TRANSCUTANEOUS BLOOD GAS MONITOR TCM4 SERIES 391-880 manufactured by Radiometer Medical Aps.

Event Text Entries

[4462349] The pt (a neonatal boy born in (b)(6) gestation week) received a burn mark on the right thigh. The temperature was set to 43 degrees celsius, and the electrode was placed on the same location for 7 hours. According to the operator's manual (tcm4/40 om page 6-12) the probe should be relocated after 4 hours at least.
Patient Sequence No: 1, Text Type: D, B5


[12082791] According to the complaint the temp of the sensor was set to 43 degrees celsius and the sensor was placed on the same location for 7 hrs. According to the operator's manual (tcm4/40 om page 6-12) the probe should be relocated after 4 hrs at least. A zero check on the sensor was performed and it passed fine. The temp was stable at 42. 9 celsius so it was not an issue with the sensor. Additionally, the field service engineer completed the tcm4 complete alignment, and monitor performance test and everything passed fine. Since no error was identified on the unit or sensor, and it was stated that the sensor was not relocated every 4 hrs as it should be this case is suspected to be caused by a user error.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3002807968-2014-00017
MDR Report Key3874109
Report Source05
Date Received2014-05-29
Date of Report2014-05-02
Date of Event2014-04-01
Date Mfgr Received2014-05-02
Device Manufacturer Date2010-01-01
Date Added to Maude2014-06-17
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer StreetAKANDEVEJ 21
Manufacturer CityBRONSHOJ 2700
Manufacturer CountryDA
Manufacturer Postal2700
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTRANSCUTANEOUS BLOOD GAS MONITOR
Generic NameTCM4 MONITORING SYS
Product CodeLKD
Date Received2014-05-29
Model NumberTCM4 SERIES
Catalog Number391-880
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerRADIOMETER MEDICAL APS
Manufacturer AddressAKANDEVEJ 21 BRONSHOJ 2700 DA 2700


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2014-05-29

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.