URINARY DRAINAGE BAGS 35 030 3 -UMUK

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,07,08 report with the FDA on 2014-05-15 for URINARY DRAINAGE BAGS 35 030 3 -UMUK manufactured by Unomedical S.r.o..

Event Text Entries

[4566622] It was reported that products were rec'd with no labels.
Patient Sequence No: 1, Text Type: D, B5


[12088597] Based on the available info, this event is deemed a reportable malfunction. A final quality investigation was performed on (b)(4) 2011. The investigation results stated the following, rec'd one box with 25 bags packed into polybags. The labels which should be placed in the middle of the polybags are missing on each of them. History records have been checked and no deviation was observed during the mfg process. No other complaints of this nature have been rec'd on the same batch and also on the same reference code. The defect occurred due to operator mistake who forgot to put the labels on the polybags during the packing process. All relevant operators will be retrained. Corrective action has been implemented. Effectiveness of action was verified by inspection after training. Corrective action was effective. No add'l pt/event details have been provided to date. Should add'l info become available, a f/u report will be submitted. This mdr is being reported as a result of a retrospective review of complaint records conducted by (b)(4) for complaints rec'd from (b)(4) 2011 - (b)(4) 2013.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3005778470-2014-00023
MDR Report Key3876086
Report Source01,07,08
Date Received2014-05-15
Date of Report2011-05-24
Date of Event2011-05-24
Date Mfgr Received2011-05-24
Date Added to Maude2014-06-30
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMATTHEW WALENCIAK, INTERIM ASSOCI
Manufacturer Street211 AMERICAN AVE
Manufacturer CityGREENSBORO NC 27409
Manufacturer CountryUS
Manufacturer Postal27409
Manufacturer Phone9083779293
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameURINARY DRAINAGE BAGS
Generic NameCOLLECTOR, URINE, (AND ACCESSORIES)
Product CodeFCN
Date Received2014-05-15
Returned To Mfg2011-06-07
Model Number35 030 3 -UMUK
Catalog Number35 030 3 -UMUK
Lot Number403778
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerUNOMEDICAL S.R.O.
Manufacturer AddressPRIEMYSELNY PARK 3 MICHALOVCE 07101 LO 07101


Patients

Patient NumberTreatmentOutcomeDate
10 2014-05-15

© 2024 FDA.report