MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2014-05-29 for TRANSCUTANEOUS BLOOD GAS MONITOR TCM4 SERIES 391-880 manufactured by Radiometer Medical Aps.
[20324047]
The patient (a neonatal boy born in (b)(6) gestation week) received a burn mark on the right thigh. The temperature was set to 43 degrees celsius, and the electrode was placed on the same location for 7 hours. According to the operator's manual (tcm4/40 om page 6-12) the probe should be relocated after 4 hours at least.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 1523456-2014-00002 |
MDR Report Key | 3879434 |
Date Received | 2014-05-29 |
Date of Report | 2014-05-02 |
Date of Event | 2014-04-01 |
Date Facility Aware | 2014-05-02 |
Report Date | 2014-05-02 |
Date Reported to FDA | 2014-05-29 |
Date Reported to Mfgr | 2014-05-28 |
Date Added to Maude | 2014-06-19 |
Event Key | 0 |
Report Source Code | Distributor report |
Manufacturer Link | N |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Reporter Occupation | OTHER HEALTH CARE PROFESSIONAL |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Street | 810 SHARON DR. |
Manufacturer City | WESTLAKE OH 441451598 |
Manufacturer Country | US |
Manufacturer Postal | 441451598 |
Manufacturer G1 | RADIOMETER AMERICA, INC. |
Manufacturer Street | 810 SHARON DR. |
Manufacturer City | WESTLAKE OH 44145159 |
Manufacturer Country | US |
Manufacturer Postal Code | 44145 1598 |
Single Use | 0 |
Previous Use Code | 0 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | TRANSCUTANEOUS BLOOD GAS MONITOR |
Generic Name | TCM4 MONITORING SYSTEM |
Product Code | LKD |
Date Received | 2014-05-29 |
Model Number | TCM4 SERIES |
Catalog Number | 391-880 |
Operator | HEALTH PROFESSIONAL |
Device Availability | Y |
Device Age | 4 YR |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | RADIOMETER MEDICAL APS |
Manufacturer Address | AKANDEVEJ 21 BRONSHOJ DK-2700 DA DK-2700 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 1. Other | 2014-05-29 |